Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Illustration 3-10: Cable Curvature
NOTE:
In case of on grade cabling, because of the minimum curvature constraint of 150 mm (6"), the
cable will have to come from the side between Innova LC Positioner and patient table.
3.3.6 Seismic areas
Chiller: The seismic kit for Lytron is included with the chiller when shipped.
In Seismic areas all cabinets must be anchored to the floor. See
Basic Innova LC System Compatibility for referential documents. C1 & C2 Cabinets: Wall
support 2285242.
Every sub–system is delivered on site with its proper seismic kit.
•
Monitor Flat Panel Seismic Kit: 2353317
•
VCIM seismic kit: 2365510.
Anti–seismic means be installed before opening the system for normal use.
Refer to Calculations for Seismic Anchoring Information.
3.4 Ceiling
Aluminum rails support the In–Room Monitor bridge used in Innova system X–ray rooms.
Reference:
For additional details on ceiling requirements for stationary rails, refer to: - Direction
46–019639, Advantx (VHLA) XT Stationary Rails Installation and Adjustment.
66
Direction 5314528-2-1EN, Revision 1
Chapter 2, Section 3,
3 Structural Requirements