Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Table 3-11: D distance
Omega IV Compact
Omega V Long
Omega V non motorized Long
Tilt Table
* in case of an upgrade, the Tilt Table is positioned at same position as the replaced Omega
NOTE:
Table.
Due to the plastic bushing used in the USA to protect cables from the sharp
edges of conduits it is necessary to place the cable conduit inside the table cable
access opening but the height of the outcoming conduit plus bushing is limited to
1/2 in (12.7 mm).
NOTE:
Refer to table Chemical anchors Pull out efforts and recommendations in
2,
Mounting Requirements for pull out effort on each fixation bolts.
3.3.5 Water Pipe Requirements
•
An Innova LC System uses a Performix 160 A X–Ray tube with a recirculating chiller.
•
Two water hoses are supplied to allow water circulation between Innova LC Positioner
and chiller.
•
In some countries, it is forbidden to run electrical cables and water pipes in the same
conduit. In this case, two separate conduits are required. But then a problem arises
at the level of the Innova LC Positioner entrance.
•
Depending on the method used, one of the two options shown hereafter (Illustrations
Illustration 3-7
Chapter 3 Physical Requirements of Room
ANGIO / CARDIO
NA
1278 mm (50.3 in)
1278 mm (50.3 in)
1278 mm (50.3 in)
and
Illustration 3-8
) must be used.
Direction 5314528-2-1EN, Revision 1
CARDIO / NEURO
1395 mm (54.9 in)
1395 mm (54.9 in)
1395 mm (54.9 in)
1278 mm (50.3 in)
Chapter 5, Section
63