2.4 Injector Mounting Requirements
Table accessory rail load consideration:
The maximum load per table accessory rail is 40 kg (88 lbs) at 150 mm (0.49 ft) (60
N.m or 44.25 ft/lbs). Therefore:
•
•
•
•
Chapter 5 Physical Characteristics
Only light extra load not exceeding 5 kg (11 lbs) at 100 mm (0.33 ft) (i.e IV
pole with its accessories, pressure head...) is authorized on the same
table accessory rail as the injector.
Never install injector and radiation protection on the same table accessory
rail.
Typical installation on the front table accessory rail is Smart handle or
Smart box, Table Side System Control (TSSC), InnovaCentral/Touchscreen,
Table panning device and cables support.
If needed an optional rail can be installed at table foot end of the Omega
V table for other options.
Direction 5314528-2-1EN, Revision 1
147