2285052 - Above Grade, Floor Anchor With Insert - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
2.3.3.4 2285052 – Above grade, Floor anchor with insert
Table 5-11:
Name
Item
Floor Anchor 46-302265P1
Grout assy
Dowel
Water
Electric
Separator
Chapter 5 Physical Characteristics
Part #
Floor Anchor Bolt; 5/8-10 x 6 in.
drilling
12 anchors for LC Positioner and
4 anchors for Table
2285055
- 10 kg Powdered Mortar Ardex K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
2290937-2
Wood Dowel; 16 mm diameter
2268647
Manufactured part introduced in
Positioner key hole to fit 1 electrical
conduit of 6".
Direction 5314528-2-1EN, Revision 1
Description
16 pc
1 kit
12 pc
1 pc
Quan.
Notes
Above Grade anchor
mounting method hardware
Used in constructing LC
Positioner grout dam
Electrical and water conduits
separated inch size.