Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
14 MIS (Master Interconnect System)
Innova system interconnect cables are described in MIS (Master Interconnect System)
documents. These documents specify all interconnections between components within the
system.
Reference: For specific Vascular system interconnect maps and connection details, refer to
the following
•
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ MIS Map
•
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ MIS Charts
General Guidelines
Innova System introduce a new system interconnect with a star distribution for all cables from
the technical area. The cables are shipped on spools to create cable groups. Cable group 1
for Exam room and cable group 2 for control room. The cable group shall be put in place
during the same action. The cables are routed in the same duct.
The HV cables could be pulled separately.
180
Direction 5314528-2-1EN, Revision 1
14 MIS (Master Interconnect System)