Injectors; Remote Injector (Rack Mount); Pedestal Injector; Injector L.f. Illumina - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

11 Injectors

The injector is provided with an additional ground cable (P/N #2135737) to meet
equipotentiality requirement at patient vicinity.

11.1 Remote Injector (rack mount)

A power cord is supplied with the injector.
If the voltage is 110 VAC, the cord should be connected inside the bulkhead positioner
cabinet, but it is highly recommended to connect it to a wall outlet fed by the main disconnect
room device.

11.2 Pedestal Injector

A power cord is supplied with the injector.
In all cases (110 VAC or 230 VAC), it will be connected to a wall outlet near the operator
location, fed by the main disconnect room device.

11.3 Injector L.F. ILLUMINA

As a result of power demand availability for all functions of this new injector, in any cases it
will be connected to a wall outlet sized to 1500 W, 110 VAC or 230 VAC as nominal voltage.
The wall outlet is fed by the main disconnect room device.
Chapter 6 Electrical Connections
Direction 5314528-2-1EN, Revision 1
177

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq

Table of Contents