Therapy System
User Manual
Operation & Installation
Instructions for:
Color Series
2765CS- Two Channel Electrotherapy System with sEMG
2755CS- Two Channel Electrotherapy System
2762CC- Two Channel Combination System with sEMG
2752CC- Two Channel Combination System
Monochromatic Series
Electromagnetic
2773MS- Two Channel Electrotherapy System
Compatibility
2772MC- Two Channel Combination System
(EMC) Tables
Optional Modules
2770- Channel 3/4 Electrotherapy Module
DJO is an ISO 13485 Certified Company

Advertisement

Table of Contents
loading

Summary of Contents for Chattanooga Intelect Advanced 2765CS

  • Page 1 Therapy System User Manual Operation & Installation Instructions for: Color Series 2765CS- Two Channel Electrotherapy System with sEMG 2755CS- Two Channel Electrotherapy System 2762CC- Two Channel Combination System with sEMG 2752CC- Two Channel Combination System Monochromatic Series Electromagnetic 2773MS- Two Channel Electrotherapy System Compatibility 2772MC- Two Channel Combination System (EMC) Tables...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Intelect® Advanced Therapy System FOREWORD ..............1 GENERAL TERMINOLOGY .
  • Page 3 TABLE OF CONTENTS Intelect® Advanced Therapy System INTELECT ADVANCED MONOCHROMATIC SERIES Applicator Coupling............39 THERAPY SYSTEMS.
  • Page 4 TABLE OF CONTENTS Intelect® Advanced Therapy System Save to Patient Data Card ..........50 Pause Treatment .
  • Page 5 TABLE OF CONTENTS Intelect® Advanced Therapy System Erasing Patient Data Card ..........72 View Waveform Rationale.
  • Page 6 TABLE OF CONTENTS Intelect® Advanced Therapy System Start Treatment ............92 Disconnect Ribbon Cable at Module.
  • Page 7: Foreword

    FOREWORD Intelect® Advanced Therapy System This manual has been written for the users of the Intelect Advanced Therapy Systems. It contains general information on the operation, precautionary practices, and maintenance information. In order to maximize use, efficiency, and the life of the system, please read this manual thoroughly and become familiar with the controls, as well as the accessories before operating the system.
  • Page 8: Precautionary Definitions

    SAFETY PRECAUTIONS Intelect® Advanced Therapy System PRECAUTIONARY DEFINITIONS The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definition of these symbols are as follows; DANGER CAUTION Caution-...
  • Page 9: Cautions

    • DO NOT operate the Intelect Advanced Therapy System when malfunction of internal components of the system and therefore connected to any unit other than Chattanooga devices. create a risk of injury to the patient. • DO NOT operate this unit in an environment where other devices •...
  • Page 10: Warnings

    SAFETY PRECAUTIONS Intelect® Advanced Therapy System WARNINGS WARNING WARNING • Powered muscle stimulators should be used only with the leads and These devices are restricted to sale by, or on the order of, a physician electrodes recommended for use by the manufacturer. or licensed practitioner.
  • Page 11 SAFETY PRECAUTIONS Intelect® Advanced Therapy System WARNINGS (continued) WARNING Stimulation should not be applied over the anterior neck or mouth. Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing.
  • Page 12: Dangers

    SAFETY PRECAUTIONS Intelect® Advanced Therapy System DANGERS DANGER • Stimulus delivered by the TENS waveforms of this device, in certain configurations, will deliver a charge of 25 microcoulombs (µC) or greater per pulse and may be sufficient to cause electrocution. Electrical current of this magnitude must not flow through the thorax because it may cause a cardiac arrhythmia.
  • Page 13: Electrotherapy Indications, Contraindications, And Adverse Effects

    SAFETY PRECAUTIONS Intelect® Advanced Therapy System ELECTROTHERAPY INDICATIONS, CONTRAINDICATIONS, AND ADVERSE EFFECTS INDICATIONS FOR VMS, VMS BURST, RUSSIAN, TENS, HIGH Contraindications VOLTAGE PULSED CURRENT (HVPC), INTERFERENTIAL AND • This device should not be used for symptomatic local pain relief PREMODULATED WAVEFORMS unless etiology is established or unless a pain syndrome has •...
  • Page 14: Additional Precautions

    SAFETY PRECAUTIONS Intelect® Advanced Therapy System ELECTROTHERAPY INDICATIONS, CONTRAINDICATIONS, AND ADVERSE EFFECTS (continued) • With TENS waveforms, isolated cases of skin irritation may ADDITIONAL PRECAUTIONS • occur at the site of electrode placement following long-term Caution should be used for patients with suspected or application.
  • Page 15: Safety Precautions

    SAFETY PRECAUTIONS Intelect® Advanced Therapy System ULTRASOUND INDICATIONS AND CONTRAINDICATIONS INDICATIONS FOR ULTRASOUND • Application of therapeutic deep heat for the treatment of • Other contraindications are patients suspected of carrying selected sub-chronic and chronic medical conditions such as: serious infectious disease and disease where it is advisable for •...
  • Page 16: Intelect Advanced Electrotherapy And Combination Therapy Systems

    NOMENCLATURE Intelect® Advanced Therapy System INTELECT ADVANCED ELECTROTHERAPY AND COMBINATION THERAPY SYSTEMS Two (2) Channel Electrotherapy System Two (2) Channel Combination System 1. Two (2) Channel Electrotherapy System 1. Two Channel Combination System 2. User Interface (See Page 2. User Interface (See Page 3.
  • Page 17: Front Access Panel

    NOMENCLATURE Intelect® Advanced Therapy System INTELECT ADVANCED ELECTROTHERAPY AND COMBINATION THERAPY SYSTEMS (continued) Front Access Panel Rear Access Panel 1. Front Access Panel Lanyard When reinstalling the Front Access Panel, make certain the 1. Screen Contrast Control (Not functional on Color Systems) Lanyard does not become kinked.
  • Page 18: User Interface

    NOMENCLATURE Intelect® Advanced Therapy System USER INTERFACE (See Page Rear Access Panel User Interface (Color Shown) Ultrasound LED Coupling Indicator (Combination only) Ultrasound Applicator- 5 cm Standard. (Optional 1 cm , 2 cm and 10 cm applicators available (Combination only) Intensity Knob Cable and Lead Wire Hook (See Page...
  • Page 19: Symbol Definitions

    NOMENCLATURE Intelect® Advanced Therapy System SYMBOL DEFINITIONS Below are the definitions for all of the symbols used in the Intelect Advanced hardware and software. Study and learn these symbols before any operation of the system. System System Hardware Symbols Optional Module and Accessory Symbols Software Symbols CONTRAST CONTROL Channel 3/4...
  • Page 20: Nomenclature

    NOMENCLATURE Intelect® Advanced Therapy System Below are the definitions for all of the unique terminology used throughout this manual. Study these and become familiar with these terms for ease of system operation and familiarization with the components and control functionality of the Intelect Adavnced Therapy page 13 System.
  • Page 21: System Specifications

    SPECIFICATIONS Intelect® Advanced Therapy System SYSTEM SPECIFICATIONS DIMENSIONS Width Combination System ......28.9 cm (11.375”) Electrotherapy System .
  • Page 22: Waveform Specifications

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS IFC (Interferential) Traditional (4 Pole) TENS- Asymmetrical Biphasic Interferential Current is a medium frequency waveform. Current The Asymmetrical Biphasic waveform has a short pulse duration. The Asymmetrical Biphasic waveform has a short pulse duration. is distributed through of two channels (four electrodes).
  • Page 23: Tens- Symmetrical Biphasic

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) TENS- Symmetrical Biphasic TENS- Alternating Rectangular The Symmetrical Biphasic waveform has a short pulse duration The Alternating Rectangular waveform is an interrupted biphasic and is capable of strong stimulation of nerve fibers in the skin current with a rectangular pulse shape.
  • Page 24: Tens- Monophasic Rectangular

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) TENS- Monophasic Rectangular High Voltage Pulsed Current (HVPC) The Monophasic Rectangular waveform is an interrupted unidirectional current with a rectangular pulse shape. This The High Voltage Pulsed Current (HVPC) has a very brief pulse waveform is commonly used with electrodiagnostic testing and duration characterized by two distinct peaks delivered at clinically to stimulate denervated muscle.
  • Page 25: Vms Tm

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) Diadynamic Waveforms The Diadynamic waveforms are rectified alternating currents. The VMS is a symmetrical biphasic waveform with a 100 µsec interphase alternating current is modified (rectified) to allow the current to interval. Because the pulse is relatively short, the waveform has a low flow in one direction only.
  • Page 26: Ifc Premodulated (2P)

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) Russian IFC Premodulated (2p) Russian Current is a sinusoidal waveform, delivered in bursts or Premodulated Current is a medium frequency waveform. Current series of pulses. This method was claimed by its author (Kots) comes out of one channel (two electrodes).
  • Page 27: Microcurrent

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) Microcurrent Burst Microcurrent is a monophasic waveform of very low intensity. The literature reports beneficial effects of this waveform in the VMS Burst is a symmetrical biphasic waveform delivered in a burst treatment of wounds.
  • Page 28: Monophasic: Monophasic Rectangular Pulsed

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) MONOPHASIC: Monophasic Rectangular Pulsed MONOPHASIC: Monophasic Triangular Pulsed The Monophasic Triangular Pulsed waveform is an interrupted The Monophasic Rectangular Pulsed waveform is an interrupted unidirectional current with a triangular pulse shape. unidirectional current with a rectangular pulse shape. Output Mode .
  • Page 29: Galvanic: Continuous

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) GALVANIC: Continuous GALVANIC: Interrupted Galvanic Current is a direct current flowing in one direction only. Galvanic Current is a direct current flowing in one direction only. The current can be continuous or interrupted. The current can be continuous or interrupted.
  • Page 30: Träbert (Ultrareiz)

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) SURGED: Monophasic Rectangular Träbert (Ultrareiz) A series of rectangular, monophasic pulses. The pulses surge to It is a monophasic waveform with a phase duration of 2 ms and a maximum power, hold and then decrease before the pause. This pause of 5 ms resulting in a frequency of approximately 143 Hz.
  • Page 31: Surged: Monophasic Triangular

    SPECIFICATIONS Intelect® Advanced Therapy System WAVEFORM SPECIFICATIONS (continued) SURGED: Monophasic Triangular A series of triangular, monophasic pulses. The pulses surge to maximum power, hold and then decrease before the pause. This waveform is well suited for muscle strengthening. Output Mode ......... .Electrodes Output Intensity .
  • Page 32: Specifications

    SPECIFICATIONS Intelect® Advanced Therapy System ULTRASOUND SPECIFICATIONS Ultrasound Frequency ......1 MHz, ± 5%; 3.3 Mhz, ±5% Effective Radiating Areas.
  • Page 33: Intelect Advanced Color Series Therapy Systems

    SET UP Intelect® Advanced Therapy System INTELECT ADVANCED COLOR SERIES THERAPY SYSTEMS Remove the Therapy System and all accessories from shipping cartons. Visually inspect for damage. Report any damage to the carrier. Color Series Color Series Standard Features Mains Power Cords Optional Accessories Order No.
  • Page 34: Intelect Advanced Monochromatic Series Therapy Systems

    SET UP Intelect® Advanced Therapy System INTELECT ADVANCED MONOCHROMATIC SERIES THERAPY SYSTEMS Remove the Therapy System and all accessories from shipping cartons. Visually inspect for damage. Report any damage to the carrier. Monochromatic Series Optional Accessories Monochromatic Series Standard Features Mains Power Cords Order No.
  • Page 35: Therapy System Set Up

    SET UP Intelect® Advanced Therapy System THERAPY SYSTEM SET UP Accessing Operator Utilities Select the row of alpha or numeric SELECT ROW AND CHARACTER BUTTONS characters desired by pushing the Plug unit into wall outlet. POWER SWITCH button beside the corresponding Turn system On.
  • Page 36: Restore Default Protocols

    SET UP Intelect® Advanced Therapy System THERAPY SYSTEM SET UP (continued) Restore Default Protocols Restore Default Unit Settings Press Restore Default Protocols Press the Restore Default Unit RESTORE DEFAULT RESTORE DEFAULT UNIT button. Settings button to restore the PROTOCOLS BUTTON SETTINGS BUTTON system defaults.
  • Page 37: Erase Patient Data Card

    SET UP Intelect® Advanced Therapy System THERAPY SYSTEM SET UP (continued) Erase Patient Data Card Set Date and Time SET DATE Install Patient Data Card to be Press Set Date and Time button. INSERT ERASE PATIENT AND TIME erased into Patient Data Card PATIENT CARD BUTTON BUTTON...
  • Page 38: Setting System Volume

    SET UP Intelect® Advanced Therapy System THERAPY SYSTEM SET UP (continued) Setting System Volume Ultrasound Coupling This warning system works in conjunction with the Applicator Press Volume button until VOLUME LED to alert the user should the Sound Head become uncoupled the desired system volume is BUTTON from the patient.
  • Page 39: Display Unit Version Information

    SET UP Intelect® Advanced Therapy System THERAPY SYSTEM SET UP (continued) Display Unit Version Information Pad Contact Quality The Pad Contact Quality feature indicates to the user the contact Press the Display Unit Version PRESS DISPLAY UNIT quality of the electrodes on the patient. This function, if On, Information button to show VERSION BUTTON TO VIEW displays a bar graph at the bottom of Treatment Review screen for...
  • Page 40: Set

    SET UP Intelect® Advanced Therapy System THERAPY SYSTEM SET UP (continued) Select Language Connecting Accessories to the Therapy System To change the language Install Lead Wires, Ultrasound Applicator, Patient Interrupt Switch, LANGUAGE displayed on the system, press and any other accessories according to the Front Access Panel as BUTTON the Language button until the page 13...
  • Page 41: Electrotherapy Patient Preparation

    PATIENT PREPARATION Intelect® Advanced Therapy System ELECTROTHERAPY PATIENT PREPARATION Electrode Placement WARNING • Examine the skin for any wounds and clean the skin. Apply the electrodes to the treatment area. • • Keep electrodes separated during treatment. Electrodes in contact with Ensure the electrodes are applied securely to the skin.
  • Page 42: Dura-Stick™ Electrodes

    ELECTROTHERAPY PATIENT PREPARATION (continued) Dura-Stick™ Electrodes Reusable Carbon Electrodes Chattanooga Dura-Stick Electrodes are a self adhesive, single If used for delivery of electrotherapy, the Carbon Electrodes must patient, one time use disposable product designed specifically for be inserted into the sponges moistened with distilled water prior to use with Chattanooga Electrotherapy systems.
  • Page 43: Dura-Stick™ Electrode Instructions

    PATIENT PREPARATION Intelect® Advanced Therapy System ELECTROTHERAPY PATIENT PREPARATION (continued) Dura-Stick™ Electrode Instructions Connecting Lead Wires Securing Electrodes Insert the lead with the Red (+) electrode Remove the Dura-Stick Electrodes from connector into one Dura-Stick Electrode. the protective backing and apply to the Insert the lead with the Black (-) electrode treatment area as prescribed.
  • Page 44: Reusable Carbon Electrodes

    PATIENT PREPARATION Intelect® Advanced Therapy System ELECTROTHERAPY PATIENT PREPARATION (continued) Reusable Carbon Electrodes Connecting Lead Wires Conductive Medium Securing Electrodes Insert the lead with the Red (+) electrode Use wet sponges or liberally apply Use the Nylatex Wrap to secure each ®...
  • Page 45: Patient Preparation

    PATIENT PREPARATION Intelect® Advanced Therapy System ULTRASOUND PATIENT PREPARATION Preparing Treatment Area Conductive Medium Examine the skin for any wounds and clean the skin. Liberally apply Conductor™ APPLY Transmission Gel or equivalent to Size of Applicator CONDUCTIVE GEL the treatment area on the patient. View the Sound Head Recommendation in the Treatment Review screen for Ultrasound (as a reference point only) prior to administering treatment.
  • Page 46: Operator Interface

    OPERATION Intelect® Advanced Therapy System OPERATOR INTERFACE The Intelect Advanced Therapy System user Operator Interface houses all of the functions and controls necessary for the operator to access all operator utilities, modalities, and parameters for modification and system set up. 6.
  • Page 47: Home Screen

    OPERATION Intelect® Advanced Therapy System HOME SCREEN The Intelect Advanced Home screen affords access to all of the system modalities and functions. The area surrounding the screen has 10 modality and parameter modification buttons. Electrotherapy Accesses all the available waveforms and parameter editing controls. Quick Link Indications Accesses specific pre-programmed indications, for general reference only, which aid in selecting the proper waveform and electrode placement for particular...
  • Page 48: Electrotherapy Screen

    OPERATION Intelect® Advanced Therapy System ELECTROTHERAPY SCREEN The screen allows the operator to access, set up, and modify parameters of each of the available waveforms within the Intelect Advanced Therapy System. The following pages give a general explanation of a treatment setup. page Refer to the Specifications section, beginning on for detailed specifications of the system and each available...
  • Page 49: General Electrotherapy Waveform Set Up

    OPERATION Intelect® Advanced Therapy System GENERAL ELECTROTHERAPY WAVEFORM SET UP The following information is an example of step by step set up for the Electrotherapy waveforms. All waveforms in the Intelect Advanced Therapy System are set up and edited in the same basic fashion. The following set up instructions use IFC Traditional (4p) Waveform. Prepare Patient Select Waveform View Waveform Description...
  • Page 50: View Electrode Placement

    OPERATION Intelect® Advanced Therapy System GENERAL ELECTROTHERAPY WAVEFORM SET UP (continued) View Electrode Placement Edit Waveform Parameters Install Patient Interrupt Switch Press the Electrode Placement button to Press Edit button to access waveform Make certain the Patient Interrupt Switch is view the most commonly used electrode parameters.
  • Page 51: Patient Interrupt Switch

    OPERATION Intelect® Advanced Therapy System GENERAL ELECTROTHERAPY WAVEFORM SET UP (continued) Patient Interrupt Switch Set Waveform Intensity Start Treatment Give Patient Interrupt Switch to patient Set intensity by rotating the Intensity Control Press the Start button to begin therapy and explain that pressing the Red button Knob to the prescribed level.
  • Page 52: Pause Treatment

    OPERATION Intelect® Advanced Therapy System GENERAL ELECTROTHERAPY WAVEFORM SET UP (continued) Save to Patient Data Card Pause Treatment Stop Treatment Stop Treatment Stop Treatment After session is complete, press the Save pages 63 to Patient Card button. Refer to Press the Pause button to pause therapy To Stop treatment, press the Stop button To Stop treatment, press the Stop button through 72...
  • Page 53: Adjusting Electrotherapy Channel Parameters During Treatment

    OPERATION Intelect® Advanced Therapy System ADJUSTING ELECTROTHERAPY CHANNEL PARAMETERS DURING TREATMENT The Electrotherapy channel parameters may be changed during a treatment session without pausing or stopping the treatment. The waveform Intensity may be increased or decreased at any time during the session without utilizing this process. Edit Channel Parameters Select Channel Press the Edit button.
  • Page 54: Ultrasound

    PARAMETER RATIONALE Use only Intelect Advanced Ultrasound SOUND HEAD BUTTON RECOMMENDATION Applicators. Previous models of BUTTON Chattanooga Ultrasound Applicators will NEXT not work with the Intelect Advanced PAGE Therapy System. BUTTON Select Modality Press the Ultrasound button on the Home screen.
  • Page 55: Edit Ultrasound Parameters

    OPERATION Intelect® Advanced Therapy System ULTRASOUND (continued) Edit Ultrasound Parameters Head Warming Set Ultrasound Intensity Press Edit button to access ultrasound The Intelect Advanced Therapy System Set intensity by rotating the Intensity Control parameters. incorporates a Head Warming feature that Knob to the prescribed level.
  • Page 56: Start Treatment

    OPERATION Intelect® Advanced Therapy System ULTRASOUND (continued) Pause Treatment Stop Treatment Stop Treatment Stop Treatment Start Treatment Press the Pause button to pause therapy To Stop treatment, press the Stop button Press the Start button to begin therapy session and maintain remaining time. To once.
  • Page 57: Editing Ultrasound From Home Screen

    OPERATION Intelect® Advanced Therapy System ADJUSTING ULTRASOUND PARAMETERS DURING TREATMENT The ultrasound parameters may be changed during a treatment session without pausing or stopping the treatment. The following information provides instructions for changing ultrasound treatment parameters during a treatment session. The ultrasound intensity may be increased or decreased at any time during the session without utilizing this process.
  • Page 58: Quick Link Indications

    OPERATION Intelect® Advanced Therapy System QUICK LINK INDICATIONS The Intelect Advanced Therapy System incorporates a unique Quick Link Indications section which allows the user to select specific Clinical Indications and apply the most common therapy for the Indication selected. All modalities are editable, in their normal editing fashion, in order to customize the treatment for each patient’s prescribed therapy.
  • Page 59: View Waveform Description

    OPERATION Intelect® Advanced Therapy System QUICK LINK INDICATIONS (continued) View Electrode Placement Edit Waveform Parameters View Waveform Description Press the Electrode Placement button to Press Edit button to access waveform Press the Waveform Description button view the most commonly used electrode parameters.
  • Page 60: Install Patient Interrupt Switch

    OPERATION Intelect® Advanced Therapy System QUICK LINK INDICATIONS (continued) Patient Interrupt Switch Install Patient Interrupt Switch Make certain the Patient Interrupt Switch is Give Patient Interrupt Switch to patient Reset intensity and press the Start button page 13 connected to the unit. Refer to and explain that pressing the Red button to resume session.
  • Page 61: Setting Waveform Intensity

    OPERATION Intelect® Advanced Therapy System QUICK LINK INDICATIONS (continued Setting Waveform Intensity Start Treatment Pause Treatment Set intensity by rotating the Intensity Control Press the Start button to begin therapy Press the Pause button to pause therapy Knob to the prescribed level. session.
  • Page 62: Stop Treatment

    OPERATION Intelect® Advanced Therapy System QUICK LINK INDICATIONS (continued) Editing Parameters during Treatment Save to Patient Data Card Stop Treatment Stop Treatment Stop Treatment Session After session is complete, press the Save The parameters may be edited during pages 63 to Patient Card button.
  • Page 63: Combination

    OPERATION Intelect® Advanced Therapy System COMBINATION The Intelect Advanced Therapy System Combination modality allows the user to select and use ultrasound therapy in combination with electrical muscle stimulation. Combination therapy utilizes the Ultrasound modality in conjuction with High Voltage Pulsed Current (HVPC), IFC (4p), IFC (2p), Asymmetrical Biphasic, Symmetrical Biphasic, or VMSTM to generate a therapeutic effect.
  • Page 64: View Electrode Placement

    OPERATION Intelect® Advanced Therapy System COMBINATION (continued) View Electrode Placement Access Combination Parameters Edit Ultrasound Parameters Press the Electrode Placement button to Press Edit button to access Combination Press the corresponding button to edit view the most commonly used electrode parameters.
  • Page 65: Select Waveform

    OPERATION Intelect® Advanced Therapy System COMBINATION (continued) Select Waveform Patient Interrupt Switch Edit Waveform Parameters Connect Patient Interrupt Switch to the Press the Select Waveform button. Press the Edit Stim button to edit the Therapy System. Give Patient Interrupt parameters of the waveform selected. Switch to patient and explain that pressing Press the corresponding button to edit SELECT WAVEFORM...
  • Page 66: Set Waveform Intensity

    OPERATION Intelect® Advanced Therapy System COMBINATION (continued) Set Waveform Intensity Set Ultrasound Intensity Start Treatment Set intensity by rotating the Intensity Control Press the Edit Ultrasound button. Press the Start button to begin therapy Knob to the prescribed level. Set Ultrasound intensity by rotating the session.
  • Page 67: Pause Treatment

    OPERATION Intelect® Advanced Therapy System COMBINATION (continued) Save to Patient Data Card Pause Treatment Stop Treatment Stop Treatment Stop Treatment After session is complete, press the Save pages 63 to Patient Card button. Refer to Press the Pause button to pause therapy To Stop treatment, press the Stop button through 72 for Patient Data Card Setup...
  • Page 68: Adjusting Combination Parameters During Treatment

    OPERATION Intelect® Advanced Therapy System ADJUSTING COMBINATION PARAMETERS DURING TREATMENT The channel parameters may be changed during a treatment session without pausing or stopping the treatment. The following information provides instructions for changing Combination Treatment Electrotherapy Channel and Ultrasound parameters during a treatment session. Edit Ultrasound Parameters Edit Waveform Parameters NOTE:...
  • Page 69: Patient Data Card Set Up Of New Card

    OPERATION Intelect® Advanced Therapy System PATIENT DATA CARD SET UP OF NEW CARD General Information The Intelect Advanced Therapy System incorporates a Patient Data Card reading and recording device that allows transfer of patient therapy data from the system to the card for reviewing patient modality and pain profile information. Information may be transferred to a PC via the optional Patient Data Management System.
  • Page 70: Enter Patient Id

    OPERATION Intelect® Advanced Therapy System PATIENT DATA CARD SET UP OF NEW CARD (continued) Enter Patient ID Select the row of alpha or numeric When the desired character is framed, press To discard entire entry, press the Back characters desired by pushing the button the Accept and Return Arrow button.
  • Page 71: Access Electrode Placement

    OPERATION Intelect® Advanced Therapy System PATIENT DATA CARD SET UP OF NEW CARD (continued) Electrode Placement Set Up Access Electrode Placement Electrode Placement Press the Pad button to select the Red (+) or The following information uses the IFC Press the Up, Down, Left and Right Arrow Black (-) electrode.
  • Page 72: Access Pain Map

    OPERATION Intelect® Advanced Therapy System PATIENT DATA CARD SET UP OF NEW CARD (continued) Access Pain Map Select Pain Type Add Pain Locations Press the Pain Map button to select Press the Pain Type button until the Press the Add button. The Add Pain Map the body area of the associated pain as desired description is displayed in the Pain window will display.
  • Page 73: Select Location Of Pain

    OPERATION Intelect® Advanced Therapy System PATIENT DATA CARD SET UP OF NEW CARD (continued) Editing Pain Locations Select Location of Pain Press the Add button to continue selecting, Press the Up and Down Arrow buttons to Press the Edit button on the Pain Map in sequence, the radiating path of the pain move the Pain Locator to the area of the window.
  • Page 74: Deleting Pain Locations

    OPERATION Intelect® Advanced Therapy System PATIENT DATA CARD SET UP OF NEW CARD (continued) Pain Scales Deleting Pain Locations Press the Delete button on the Pain Map The Intelect Advanced Therapy System incorporates two internationally recognized Pain window Scales; VAS Visual Analog Scale, Scale has no indicator marks or Numeric (indicated 1 through 10) for use in describing the amount of pain the patient is experiencing.
  • Page 75: Save To Patient Data Card

    OPERATION Intelect® Advanced Therapy System PATIENT DATA CARD SET UP OF NEW CARD (continued) Save to Patient Data Card the Home button. After all desired session data has been entered in the Patient Data Card screens, Remove the Patient Data Card for filing with press the Save to Patient Card button.
  • Page 76: Existing Patient Data Card Use

    OPERATION Intelect® Advanced Therapy System EXISTING PATIENT DATA CARD USE Insert Existing Patient Data Card Access Patient Data Card View Patient Data Card Insert the Patient Data Card assigned to the Press the Patient Card button to access Press the corresponding button beside the patient receiving treatment into the system Patient Data Card.
  • Page 77: Starting A New Treatment From Patient Data Card

    OPERATION Intelect® Advanced Therapy System EXISTING PATIENT DATA CARD USE (continued) Starting a New Treatment from Patient Patient Interrupt Switch Set Intensity Data Card Connect Patient Interrupt Switch to the Set intensity by rotating the Intensity Control Refer to pages 35-38 for Electrotherapy Therapy System.
  • Page 78: Start Treatment

    OPERATION Intelect® Advanced Therapy System EXISTING PATIENT DATA CARD USE (continued) Start Treatment Pause Treatment Stop Treatment Stop Treatment Stop Treatment Press the Start button to begin therapy Press the Pause button to pause therapy To Stop treatment, press the Stop button session.
  • Page 79: Clinical Resources Library Clinical Protocols

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CLINICAL PROTOCOLS™ The Intelect Advanced Clinical Resources Library contains Clinical Protocols™, User Protocols, Sequencing functions and access to the Multimedia Card (MMC) which contains the Anatomical and Pathological Libraries. Clinical Protocols™ This library is a series of protocol presets where the Body Area, Clinical Indication, Pathological Condition, and Pathological Severity are selected by the user, and the Clinical Protocols algorithm will select the parameter settings.
  • Page 80: Select Clinical Indication

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CLINICAL PROTOCOLS™ (continued) Select Clinical Indication Select Pathological Condition Select Pathological Severity Press the button beside the Clinical Press the button beside the desired Press the button beside the desired Indication desired in either the Pathological Condition.
  • Page 81: View Waveform Rationale

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CLINICAL PROTOCOLS™ (continued) View Waveform Rationale View Electrode Placement Prepare Patient pages 35 through 38 Press the Electrode Placement button to Refer to Press the Waveform Rationale button page 39 (Electrotherapy Modalities) or the Parameter view the specific electrode placement for Electrotherapy and Ultrasound...
  • Page 82: Patient Interrupt Switch

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CLINICAL PROTOCOLS™ (continued) Patient Interrupt Switch Set Modality Intensity Make certain the Patient Interrupt Switch, Set intensity by rotating the Intensity Control Give Patient Interrupt Switch to patient for the channel(s) being used, is connected Knob to the prescribed level.
  • Page 83: Start Treatment

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CLINICAL PROTOCOLS™ (continued) Pause Treatment Stop Treatment Stop Treatment Stop Treatment Start Treatment Press the Start button to begin therapy Press the Pause button to pause therapy To Stop treatment, press the Stop button session.
  • Page 84: Clinical Resources Library Creating User Protocols

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CREATING USER PROTOCOLS General Information This library is a series of protocols created by the user and stored in the system memory. The following information gives general instructions as to setting up, saving and accessing User Protocols. Should the Default Protocols be restored, through the User Utilities, all User Protocols will be permanently removed from the system.
  • Page 85: Enter User Protocol Name

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CREATING USER PROTOCOLS (continued) Enter User Protocol Name Select the row of alpha or numeric When the desired character is framed, press To discard entire entry, press the Back characters desired by pushing the button the Accept and Return Arrow button.
  • Page 86: Clinical Resources Library Deleting User Protocols

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY DELETING USER PROTOCOLS General Information The following information provides instructions for the deletion of one User Protocol at a time. Once any single User Protocol is deleted, it Deleting User Protocols cannot be recovered. Should the Default Protocols be restored, through the User Utilities, all User Protocols will be permanently removed from the system.
  • Page 87: View Waveform Rationale

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY USING USER PROTOCOLS Access User Protocols Select User Protocol View Waveform Rationale Press the Clinical Resources Library button. Press the Waveform Rationale button Press the UP and Down Arrow buttons until the Press the User Protocols button.
  • Page 88: View Electrode Placement

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY USING USER PROTOCOLS (continued) View Electrode Placement Prepare Patient Patient Interrupt Switch pages 35 through 38 Press the Electrode Placement button to Refer to Make certain the Patient Interrupt Switch, page 39 view the electrode placement for the User Electrotherapy and for Ultrasound...
  • Page 89: Set Modality Intensity

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY USING USER PROTOCOLS (continued) Set Modality Intensity Start Treatment Set intensity by rotating the Intensity Control Give Patient Interrupt Switch to patient Press the Start button to begin therapy Knob to the prescribed level. and explain that pressing the Red button session.
  • Page 90: Pause Treatment

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY USING USER PROTOCOLS (continued) Save to Patient Data Card Pause Treatment Stop Treatment Stop Treatment Stop Treatment After session is complete, press the Save pages 63 to Patient Card button. Refer to Press the Pause button to pause therapy To Stop treatment, press the Stop button through 72...
  • Page 91: Clinical Resources Library Creating New Sequences

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CREATING NEW SEQUENCES General Information This Library is a series of Electrotherapy Waveform/Current Sequences created by the user for special electrotherapy treatment purposes and stored in the system memory for recall and use. The following information gives general instructions for setting up, saving and accessing of sequences.
  • Page 92: Edit First Waveform Or Current

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CREATING NEW SEQUENCES (continued) Edit First Waveform or Current Select Second Waveform or Current Saving New Sequence Press the Edit button on the Sequence Press the Down Arrow button on the After all waveforms/currents have been screen.
  • Page 93: Enter Sequence Name

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY CREATING NEW SEQUENCES (continued) Enter Sequence Name Select the row of alpha or numeric When the desired character is framed, press To discard entire entry, press the Back characters desired by pushing the button the Accept and Return Arrow button.
  • Page 94: Clinical Resources Library Deleting Sequences

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY DELETING SEQUENCES General Information The following information provides instructions for the deletion of one user defined sequence at a time. Once any single sequence is deleted, it cannot be recovered. Should the Default Protocols be restored, through the User Utilities, all user defined sequences will be permanently removed from the system.
  • Page 95: Clinical Resources Library Using Sequences

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY USING SEQUENCES Access Sequencing Select Sequence Select Waveform/Current Press the Clinical Resources Library button. Press the Down Arrow button, on the Press the UP and Down Arrow buttons until the Press the Sequencing button. prescribed Sequence is highlighted.
  • Page 96: View Waveform Rationale

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY USING SEQUENCES (continued) View Waveform Rationale View Electrode Placement Prepare Patient pages 35 through 38 Press the Electrode Placement button Refer to Press the Waveform Description button to view the text explaining the rationale for to view the electrode placement for the Electrotherapy patient preparation Sequence selected.
  • Page 97: Set Sequence Intensity

    OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY USING SEQUENCES (continued) Set Sequence Intensity The first waveform/current in the Sequence Press the Down Arrow button until the Give Patient Interrupt Switch to patient should be highlighted. Set intensity by second waveform/current in the sequence is and explain that pressing the Red button rotating the Intensity Control Knob to the Highlighted.
  • Page 98 OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY USING SEQUENCES (continued) Pause Treatment Stop Treatment Stop Treatment Stop Treatment Start Treatment Press the Pause button to pause therapy Press the Start button to begin therapy To Stop treatment, press the Stop button session and maintain remaining time.
  • Page 99 OPERATION OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY MMC GRAPHICAL LIBRARY General Information The Clinical Resources Library contains the unique Anatomical and Pathological Graphic Libraries* from DJO. These Graphic Libraries are contained on a single Multimedia Card (MMC) and are designed to aid the operator in visually understanding and locating specific muscle groups and commonly found problems associated with Pathological Conditions as well as providing an educational tool for the clinician to use with the patient.
  • Page 100: Operation

    OPERATION OPERATION Intelect® Advanced Therapy System CLINICAL RESOURCES LIBRARY MMC GRAPHICAL LIBRARY (continued) Select Library Type Anatomical Example- Muscles Pathological Example- Rotator Cuff Tear Select the desired graphic by pressing the corresponding button. Superficial Left Side buttons- Anatomical Selections Right Side buttons- Pathological Selections PATHOLOGICAL ANATOMICAL...
  • Page 101 INSTALLATION/REMOVAL Intelect® Advanced Therapy System INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE General Information The Intelect Advanced Therapy System Channel 3/4 Electrotherapy Module is a two channel electrotherapy module intended to upgrade the Intelect Advanced Therapy System Two Channel Electrotherapy and Two Channel Combination Therapy Systems to Four Channel Electrotherapy or Combination Therapy Systems.
  • Page 102 INSTALLATION/REMOVAL Intelect® Advanced Therapy System INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Nomenclature 1. Channel 3/4 Electrotherapy Module 2. Extended Front Access Panel 3. Module to System Mounting Holes 4. Module to System Feet Alignment Indents 5. Power Cord Routing Port 6.
  • Page 103 INSTALLATION/REMOVAL Intelect® Advanced Therapy System INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Specifications DIMENSIONS Width ..........21 cm (8.250") Depth .
  • Page 104 INSTALLATION/REMOVAL Intelect® Advanced Therapy System INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Disconnect Mains Power Remove Therapy System from Cart WARNING Remove the Therapy System from the Therapy System Cart, if equipped. Refer to the Therapy System Cart User Manual for proper instructions.
  • Page 105 INSTALLATION/REMOVAL Intelect® Advanced Therapy System INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Release Ribbon Cable Position Therapy System and Module Connect Ribbon Cable Carefully align the Ribbon Cable Connector Position Therapy System and the Channel Remove and discard the vinyl label holding to the Module Connector Pins and press 3/4 Electrotherapy Module as shown.
  • Page 106 INSTALLATION/REMOVAL Intelect® Advanced Therapy System INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Front Access Panel CAUTION With a #1 Phillips Screwdriver, remove the screw retaining the existing Front Access Do not twist Ribbon Cable. If power is applied to the system with Panel.
  • Page 107 INSTALLATION/REMOVAL Intelect® Advanced Therapy System INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Install Lead Wires and Accessories Install Front Access Panel Connect Mains Power Install Lead Wires and additional accessories Install the new Extended Front Access Panel Connect the Mains Power Cord to the page 13 to Front Panel.
  • Page 108 INSTALLATION/REMOVAL Intelect® Advanced Therapy System INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Turn Therapy System On WARNING Turn the System On using the On/Off Switch. The System will automatically recognize the added Module and display a Verify that the Module installed is the Module displayed in the configuration change message.
  • Page 109 INSTALLATION/REMOVAL Intelect® Advanced Therapy System REMOVAL CHANNEL 3/4 ELECTROTHERAPY MODULE NOTE: Disconnect Mains Power It is not necessary to remove the WARNING sEMG Module from the Channel 3/4 Electrotherapy Module, if installed. Disconnect the system from the power source (outlet or remove battery module if installed) before attempting any maintenance, Remove Lead Wires and Accessories installation, removal, or replacement procedures to prevent electrical...
  • Page 110 INSTALLATION/REMOVAL Intelect® Advanced Therapy System REMOVAL CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Remove Screws Securing Module Disconnect Ribbon Cable at Module Store and Secure Ribbon Cable With a #1 Phillips Screwdriver, remove the Separate the Module from the Therapy Roll the Ribbon Cable up and store in four 4 mm x 20 mm screws securing the System and disconnect the Ribbon Cable the cavity of the Therapy System.
  • Page 111 INSTALLATION/REMOVAL Intelect® Advanced Therapy System REMOVAL CHANNEL 3/4 ELECTROTHERAPY MODULE (continued) Front Access Panel Install Lead Wires and Accessories Connect Mains Power Connect the Mains Power Cord to the With a #1 Phillips Screwdriver, remove the Re-install Lead Wires and Accessories to the Therapy System.
  • Page 112 INSTALLATION/REMOVAL Intelect® Advanced Therapy System CHANNEL 3/4 ELECTROTHERAPY MODULE REMOVAL (continued) Turn Therapy System On WARNING Turn the System On using the On/Off Switch. The System will automatically Verify that the Module installed is the Module displayed in recognize the Module has been removed the message BEFORE pressing the START button.
  • Page 113 TROUBLESHOOTING Intelect® Advanced Therapy System ERROR CODES General Information The Intelect Advanced Therapy Systems incorporate error messages, and warnings to inform the user of problems or potential problems with the system, modality, or accessories. These are numbered so the user can possibly correct the problem without the aid of service personnel.
  • Page 114 TROUBLESHOOTING Intelect® Advanced Therapy System ERROR CODES (continued) Code Type Probable Cause Possible Remedies Number Message Warning Attempting to save additional User Protocols or Sequences after Delete some User Protocols or Sequences. Refer to appropriate section of the Therapy System User system memory has reached the maximum allowed (200).
  • Page 115 TROUBLESHOOTING Intelect® Advanced Therapy System ERROR CODES (continued) Code Type Probable Cause Possible Remedies Number Message Warning Attempting to save additional User Protocols or Sequences after Delete some User Protocols or Sequences. Refer to appropriate section of the Therapy System User system memory has reached the maximum allowed (200).
  • Page 116 TROUBLESHOOTING Intelect® Advanced Therapy System ERROR CODES (continued) Code Type Probable Cause Possible Remedies Number Message Warning sEMG Data Card full. sEMG Data Card faulty. Insert a known good sEMG Data Card. If problem continues, contact dealer or DJO for service. A.
  • Page 117 TROUBLESHOOTING Intelect® Advanced Therapy System ERROR CODES (continued) Code Type Probable Cause Possible Remedies Number Message Warning Control Board Software upgrade warning. Upgrade Control Board Software to latest version. Contact dealer or DJO for latest software upgrade and instructions. Warning Stim Board Main Software upgrade warning.
  • Page 118 TROUBLESHOOTING Intelect® Advanced Therapy System ERROR CODES (continued) Code Type Probable Cause Possible Remedies Number Message Warning Wrong Laser Applicator connected to system for the protocol A. Connect correct Laser Applicator to the system. selected. B. If Applicator is connected, reset Therapy System by turning main power switch Off and On. C.
  • Page 119 MAINTENANCE Intelect® Advanced Therapy System CARING FOR THE THERAPY SYSTEM SERVICE Should the unit require service, warranty, or repair, please contact Cleaning the Therapy System the selling dealer or your local DJO customer service. With the system disconnected from the power source, clean the system with a clean, lint free cloth moistened with water and mild antibacterial soap.
  • Page 120 WARRANTY Intelect® Advanced Therapy System DJO, LLC ("Company") warrants that the Intelect Advanced Therapy System ("Product") is free of defects in material and workmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If this Product fails to function during the two year warranty period due to a defect in material or workmanship, Company or the selling dealer will repair or replace this Product without charge within a period of thirty (30) days from the date on which the Product is returned to the Company or the dealer.
  • Page 121 DJO is an ISO 13485 Certified Company DJO France SAS Centre Europeen de Fret 3 rue de Bethar 64990 Mouguerre, France T: + 33 (0) 5 59 52 86 90 F: + 33 (0) 5 59 52 86 91 djoglobal.eu/fr_FR ©...