TROUBLESHOOTING
ERROR CODES (continued)
Code
Type
Number
Message
108
Warning
109
Warning
110
Warning
111
Warning
112
Warning
113
Warning
114
Warning
115
Warning
116
Warning
117
Warning
Probable Cause
Attempting to save additional User Protocols or Sequences after
system memory has reached the maximum allowed (200).
Attempting to access protocols or sequences and none are found
in the system.
Ultrasound Applicator disconnected from system during treatment
session.
Attempting to perform Ultrasound treatment with no Applicator
connected to the system.
Ultrasound Applicator is not calibrated.
Ultrasound Applicator is too hot.
1. No Patient Data Card is inserted into the system.
2. Attempted to use an Invalid Patient Data Card.
Delete some User Protocols or Sequences. Refer to appropriate section of the Therapy System User
Manual for instructions.
A. User Protocols- No protocols have been saved in the system. Refer to Therapy System User Manual
to save User Protocols.
B. Sequences- No User Sequences have been saved in the system. Refer to Therapy System User
Manual to save Sequences.
A. Connect Ultrasound Applicator to system.
B. If Ultrasound Applicator is connected, reset system by turning power switch Off and On.
C. If problem persists, connect a known good Ultrasound Applicator. If problem continues, contact
dealer or DJO for service.
A. Connect the desired Ultrasound Applicator to the system.
B. If Ultrasound Applicator is connected, reset system by turning power switch Off and On.
C. If problem persists, connect a known good Ultrasound Applicator. If problem continues, contact
dealer or DJO for service.
Attempt to use a known good Applicator. If problem continues, contact dealer or DJO for service.
Allow Ultrasound Applicator Sound Head to cool to ambient temperature.
A. Properly insert the Patient Data Card into the system port. Refer to Therapy System User Manual for
new and existing Patient Data Card instructions.
B. Attempt to use a known good Patient Data Card.
C. Make certain a Patient Data Card and not an sEMG Data Card is being used.
D. If problem continues, contact dealer or DJO for service.
108
Intelect® Advanced Therapy System