  • Page 2: Table Of Contents

    CONTENTS FOREWORD POWERCORD INTENDED USER PROFILE STIM SET INCLUDES INTENDED ENVIRONMENT FOR USE US SET INCLUDES INTENDED USE ULTRASOUND APPLICATOR PRECAUTIONARY INSTRUCTIONS HEAD TO CART FIXATION GENERAL TERMINOLOGY CONNECTING CABLES AND INSERTING PLUGS SYSTEM SOFTWARE SYMBOLS INITIAL RECEIPT DESCRIPTION OF DEVICE MARKINGS IF UNIT SUPPLIED WITH OPTIONAL BATTERY VACUUM MODULE MARKINGS IFU DOWNLOAD...
  • Page 3 CONTENTS DEVICE USER INTERFACE WARRANTY REPAIR/OUT OF WARRANTY REPAIR WARRANTY SCREEN DESCRIPTION ELECTROMAGNETIC COMPATIBILITY (EMC) SETTINGS ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES PRINT SCREEN FUNCTION HOME SCREEN TREATMENT REVIEW SCREEN GUIDELINES SCREEN ULTRASOUND OPERATION SPS (SUGGESTED PARAMETER SETUP) TREATMENT DATA CUSTOM PROTOCOLS SHORT CUTS UNASSIGN SHORT CUT CLINICAL RESOURCES...
  • Page 4: Foreword

    INTRODUCTION FOREWORD INTENDED USE The Intelect Mobile 2 device will be used to deliver a variety This manual is intended for users of Intelect Mobile 2 . It ® of modalities to the patient, Ultrasound and electrical contains general information on operation, precautionary stimulation delivered either as stand alone therapies or in practices, and maintenance.
  • Page 5: General Terminology

    INTRODUCTION GENERAL TERMINOLOGY The following are definitions for the terminology used throughout this manual . Study these terms to become familiar with them for ease of system operation and control functionality of the Intelect Mobile 2 . ® SYSTEM SOFTWARE SYMBOLS Home Run again Back to previous screen...
  • Page 6: Description Of Device Markings

    INTRODUCTION DESCRIPTION OF DEVICE MARKINGS The markings on the unit are assurance of its conformity to the highest applicable standards of medical equipment safety and electromagnetic compatibility and conform to ISO 7010 and ISO15-223-1 One or more of the following markings may appear on the device: Refer to Instructional Manual Booklet Atmospheric Pressure Range...
  • Page 7: Ultrasound Indications

    INDICATIONS FOR USE ULTRASOUND INDICATIONS INDICATIONS ADDITIONAL PRECAUTIONS • Relief of pain from muscle spasm Additional precautions should be used when ultrasound is • Relief of pain from joint contracture used on patients with the following conditions: • Relief of pain associated with ligament sprains, •...
  • Page 8: Product Description

    DEVICE DESCRIPTION PRODUCT DESCRIPTION BATTERY MODULE (optional) The Intelect Mobile 2 ULTRASOUND is an ultrasound ® therapy system used with or without an optional Cart, allowing for the inclusion of a Vacuum module. This equipment is to be used only under the prescription and supervision of a licensed medical practitioner.
  • Page 9: Operator Interface

    DEVICE DESCRIPTION OPERATOR INTERFACE The Intelect Mobile 2 Operator Interface contains all ON/OFF switch (only active when connected to the ® the functions and controls necessary for operator access mains) to all operator utilities, modalities, and parameters for Ultrasound Applicator holder, left and right sides modification and system set up.
  • Page 10 GENERAL WARNINGS AND PRECAUTIONS CAUTION • This unit should be operated at +5°C to +40°C and 15% to 90% Relative Humidity. The unit should be transported and stored at -20°C to +60°C and 10% to 90% Relative Humidity. • Use of parts or materials other than DJO’s can degrade minimum safety. •...
  • Page 11 GENERAL WARNINGS AND PRECAUTIONS WARNING • This device should be used only under the continued supervision of a physician or licensed practitioner. • Contaminated sponges, electrodes, leadwires, and gel can lead to infection. • DO NOT operate the Intelect® Mobile 2 within the vicinity or environment of an ultrasonic diathermy system. •...
  • Page 12 GENERAL WARNINGS AND PRECAUTIONS DANGER • DO NOT connect the unit to an electrical supply without first verifying that the power supply is the correct voltage. Incorrect voltage may cause unit damage, malfunction, electrical shock, fire, or personal injury. Your unit was constructed to operate only on the electrical voltage specified on the Voltage Rating and Serial Number Plate.
  • Page 13: Intelect Mobile 2 Set Components

    15-0133 INTELECT MOBILE 2 COMBO below. 79967 Carbon electrodes 70010 STIM lead wires 15-1200 Intelect Mobile 2 Ultrasound INTL Set EU Plug 6522055 Chattanooga straps 42198 Electrodes gel 15-1201 Intelect Mobile 2 Ultrasound INTL Set All Plug 15-0144/46/47 Power cord...
  • Page 14: Ultrasound Applicator

    DETAIL DEVICE DESCRIPTION ULTRASOUND APPLICATOR 1. Applicator Head The component of the applicator that makes contact with the patient during Ultrasound or Combination therapy. 2. Applicator The assembly that connects to the system and incorporates the Applicator head. 3. LED The component of the applicator that indicates if the Applicator is coupled or uncoupled on the treatment area.
  • Page 15: Head To Cart Fixation

    SETUP INSTRUCTIONS HEAD TO CART FIXATION The optional Therapy System Cart, is designed for use with the Intelect Mobile 2 only and allows the user to easily ® transport the System from patient to patient within the clinic as well as store all necessary accessories, supplies, and applicators used for the various modalities of the System.
  • Page 16: Connecting Cables And Inserting Plugs

    5. Replace the 2 screws to close the battery cover 3. Click on "Continue" button to go to home screen IFU DOWNLOAD 1. Go to the Chattanooga website www.chattanoogarehab.com 2. Go to Intelect Mobile 2 product tab 3. Complete the registration form to be informed about new product software version availability and IFU updates.
  • Page 17: Device Connected To The Mains

    SETUP INSTRUCTIONS DEVICE CONNECTED TO THE MAINS STOP TREATMENT AND TURN OFF THE DEVICE 1. Plug the Power cord into the back of device. Plug the other Press Play/pause button to pause treatment then press stop end of the cord into an electrical outlet. on touch screen.
  • Page 18: Device Light Indicators

    SETUP INSTRUCTIONS DEVICE LIGHT INDICATORS Intelect Mobile 2 ULTRASOUND has several light indicators: FRONT PANEL INDICATORS: 1. Colors: • Light blue around Ultrasound therapy channel Left and Right 2. Behaviour: • Steady when modality is selected and output is not active •...
  • Page 19: System Specifications And Dimensions

    SYSTEM SYSTEM SPECIFICATIONS AND DIMENSIONS Width Depth Height Weight (no battery) Intelect Mobile 2 Head Unit UltraSound 34cm 35.5cm 15cm 2.8kg Cart configurations Cart (Safe working load 6.5kg) 48cm (MAX) 52cm (MAX) 96cm 10.1kg Device on cart 111 cm POWER ULTRASOUND SPECIFICATIONS Input 100 - 240 V AC, 1.0 to 0.42 A, 50/60 Hz...
  • Page 20: General System Operating And Storage Temperature

    SYSTEM IPXX Rating for Unit Unless otherwise specified, ultrasound controls accuracy is: ± 20 %. Rated to IP21 IP2* Protection against fingers or other object not greater Peak to Average Ratio: 1:1, at 50% Duty Cycle than 80mm in length and 12mm in diameter 4:1, at 20% Duty Cycle *1 Protection from vertically dripping water 9:1, at 10% Duty Cycle...
  • Page 21: Ultrasound Patient Preparation

    PATIENT PREPARATION ULTRASOUND PATIENT PREPARATION If US Coupling is “On”, the Applicator is properly coupled to the patient and administering ultrasound when the Examine the skin for any wounds and clean the skin LED is constantly illuminated. If the applicator head View the Applicator recommendation in the treatment becomes uncoupled the LED on the head will flash.
  • Page 22: Device User Interface

    USER INTERFACE DEVICE USER INTERFACE SCREEN DESCRIPTION 1. Menu Bar 2. Main Area 3. Channel Bar Each screen contains the following areas: Menu Bar Located at the top of each screen and lists the current screen name. Battery level (Black connected, Grey inacctive) Settings INTELECT MOBILE 2 ULTRASOUND USER MANUAL...
  • Page 23 USER INTERFACE Main area Located under the menu bar, this area displays icons unique to the current screen. 1. Menu Bar 2. Main Area 3. Channel Bar INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 24 USER INTERFACE Channel Bar Located at the bottom of each screen, this area displays the status information about each channel. When starting a treatment, channels are automatically assigned to the next available channel. Manual selection is done by touching the desired channel. Ultrasound status box Stim Ch 1 status box Stim Ch 2 status box...
  • Page 25: Settings

    USER INTERFACE SETTINGS The settings icon on the top right hand corner of the home screen menu bar (see page 34) offers users the opportunity to set preferences and can be accessed by pressing the '' button. Swipe vertically to see more settings On the home screen, the "current screen name"...
  • Page 26: Print Screen Function

    USER INTERFACE PRINT SCREEN FUNCTION The Intelect Mobile 2 device has a built in function allowing the user to print screen fro example to print a treatment session this performed by: » 1. insert USB drive into the USB port on the back of the Mobile 2 device »...
  • Page 27: Treatment Review Screen

    USER INTERFACE TREATMENT REVIEW SCREEN The Intelect Mobile 2 Treatment Review screens for Ultrasound includes the following information: ® 1. Touch to activate Electrode Placement 2. Adjust with Guidelines Adjustment dial: • Clockwise – Save/Overwrite default Increase settings or Custom Protocols •...
  • Page 28: Guidelines Screen

    USER INTERFACE GUIDELINES SCREEN The Guidelines for ultrasound provide the following information: Instructions for optimal US applicator use at the left side of the screen. Images illustrating US treatment area and recommended applicator choice at the right side of the screen. Touch image to view in full screen mode Swipe horizontally...
  • Page 29: Ultrasound Operation

    1. To prepare the patient’s skin for Ultrasound Therapy, prepare patient as described in the ULTRASOUND PATIENT PREPARATION section on page ..NOTE: Use only Intelect Mobile 2 Ultrasound Applicators . Previous models of Chattanooga Ultrasound Applicators will not work with the Intelect ® ® Mobile 2.
  • Page 30 USER INTERFACE 4. START TREATMENT Press the START button to begin the therapy 5. PAUSE TREATMENT Press the Start/Pause button INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 31 USER INTERFACE To resume treatment, press the Start /Pause button again Note: Pause applies to the selected channel only 6. STOP TREATMENT - First pause treatment by pressing the Start/Pause button - Then press the ‘Stop treatment ‘ box on the treatment review screen. INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 32 USER INTERFACE Note: A running treatment can only be stopped from the Pause status When treatment has completed, the Treatment Summary screen will appear with the following options: • Save » the treatment protocol to the Custom Protocols (cfr. Page..) •...
  • Page 33: Sps (Suggested Parameter Setup)

    USER INTERFACE SPS (SUGGESTED PARAMETER SETUP) The Intelect Mobile 2 has a Suggested Parameter Setup (SPS) icon that is a series of protocol presets where the body area, ® clinical indication, pathological condition and severity are selected by the user, and the suggested algorithm will select the parameter settings.
  • Page 34 USER INTERFACE 3. Select INDICATION Touch to select Swipe vertically to see more 4. Select specification Touch to select INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 35 USER INTERFACE 5. Select MODALITY/WAVEFORM Touch to select Selection refinement summary 6. SET UP TREATMENT On the treatment review screen the suggested treatment settings are displayed and you can adjust parameters to desired level. Refer to page … for detailed description of the Treatment Review Screen. Note: Never start with intensity adjustment –...
  • Page 36 USER INTERFACE 7. START TREATMENT Press the START button INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 37: Treatment Data

    USER INTERFACE TREATMENT DATA After a treatment has been completed, Treatment data can be saved on the Intelect Mobile 2 for later use on the unit. SAVE TREATMENT DATA Click on Assign To button. Treatment data can be assigned to a folder at any time of the treatment (set up, running or completed) but data will only be saved once the treatment is finished and channel is free for next treatment (after pressing EXIT button on Treatment Summary screen) The TREATMENT DATA screen appears...
  • Page 38 USER INTERFACE SAVE TREATMENT DATA TO A NEW ID: Enter ID and Save VIEW AND MANAGE TREATMENT DATA Press the TREATMENT DATA ICON on the home screen INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 39 USER INTERFACE 1. VIEW Treatment Data Select desired ID folder The TREATMENT HISTORY is displayed including all previously saved treatment sessions ranked chronologically Select session to view Treatment Summary for that session. INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 40 USER INTERFACE 2. DELETE Treatment Data Delete all IDs Delete one ID INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 41 USER INTERFACE Delete all treatment sessions Delete one session INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 42: Custom Protocols

    USER INTERFACE CUSTOM PROTOCOLS The Intelect Mobile 2 allows for a maximum of 25 custom protocols to be defined. ® SAVE A CUSTOMIZED PROTOCOL A new custom protocol may be saved at any time using SAVE button 1. Touch SAVE on the TREATMENT REVIEW or TREATMENT SUMMARY screen 2.
  • Page 43 USER INTERFACE 3. The Custom Protocol library is displayed where you can save the protocol as NEW custom protocol or OVERWRITE an existing custom protocol Create new custom protocol Select and overwrite existing protocol Alphabetical list Swipe to see more CREATE NEW CUSTOM PROTOCOL: Enter Custom Protocol Name and Save INTELECT...
  • Page 44 USER INTERFACE VIEW AND MANAGE CUSTOM PROTOCOLS Touch the CUSTOM PROTOCOLS icon on the Home Screen 1. VIEW Custom Protocol Select desired Custom Protocol INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 45 USER INTERFACE The TREATMENT REVIEW SCREEN is displayed showing the protocol settings. Start treatment or perform other actions as described in the Ultrasound Operations section 2. DELETE custom protocol Delete all protocols INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 46: Short Cuts

    USER INTERFACE SHORT CUTS Intelect Mobile 2 allows for 12 custom protocol shortcut assignments on the home screen. ASSIGN SHORTCUT Complete the following steps to assign a home screen shortcut. Unassigned Shortcut icons appear grey in colour: Press one of the unassigned “Shortcut”...
  • Page 47 USER INTERFACE Shortcut assigned on Home screen Once assigned the shortcut icon becomes the colour associated with the modality it contains INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 48: Unassign Short Cut

    USER INTERFACE UNASSIGN SHORT CUT Complete the following steps to unassign a Home screen shortcut for a customized protocol: From the Home screen, press and hold the shortcut icon you wish to unassign. The unit will display a text box asking, “Remove “My Custom Protocol 1” shortcut?” Select “No”...
  • Page 49: Clinical Resources

    USER INTERFACE CLINICAL RESOURCES The Intelect Mobile 2 contains a unique Clinical Resources Library . ® The anatomical and pathological image library are designed to aid the operator in visually understanding and locating specific muscle groups and commonly found problems associated with pathological conditions, as well as providing an educational tool for the clinician to use with the patient.
  • Page 50 USER INTERFACE 2. Touch the body part for which you wish to view information. Choose either anterior (on left of screen) or posterior (on right of screen). 3. The available images for the selected body part are displayed. Touch the image you want to see in full screen mode. INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 51 USER INTERFACE 4. Full screen image Close full screen mode INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 52: Modality/Waveform Descriptions

    USER INTERFACE MODALITY/WAVEFORM DESCRIPTIONS Complete the following steps to view the ultrasound or waveform descriptions: 1. Press the Ultrasound description icon on the Clinical Resources screen 2. Select the desired waveform (in case of Electrotherapy Waveform description) INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 53 USER INTERFACE 3. The modality or waveform description is displayed INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 54: Troubleshooting

    TROUBLESHOOTING TROUBLESHOOTING 1. All system messages, warning messages and fault messages that are generated by the device are self-explanatory excepting system error. 2. If System error occurs, note error code and contact DJO selling dealer or DJO Service Department. INTELECT MOBILE 2 ULTRASOUND USER MANUAL ®...
  • Page 55: General Accessories

    TROUBLESHOOTING GENERAL ACCESSORIES Model Number Description 15-1136 Mobile 2 Cart BATTERY (Not available at launch) Model Number Description 14-1086 Battery ULTRASOUND APPLICATORS AND GEL Model Number Description 15-0140 G16 Ultrasound Applicator 1 cm 15-0141 G16 Ultrasound Applicator 2 cm 15-0142 G16 Ultrasound Applicator 5 cm 4248 Conductor...
  • Page 56: Cleaning The Intelect Mobile

    MAINTENANCE CLEANING THE INTELECT MOBILE 2 ® With the system disconnected from the power source, clean the system with a clean, lint-free cloth moistened with water and mild antibacterial soap. If a more sterile cleaning is needed, use a cloth moistened with an antimicrobial cleaner.
  • Page 57: Calibration Requirements

    Plug the new battery to the battery connector Insert the battery in its location Replace the battery door with the 2 screws INSTRUCTION FOR SOFTWARE UPGRADE Go to the Chattanooga website REPLACEMENT BATTERY www.chattanoogarehab.com Go to Intelect Mobile 2 product tab...
  • Page 58: Warranty Repair/Out Of Warranty Repair

    SERVICE AND WARRANTY WARRANTY REPAIR/OUT OF WARRANTY This Warranty Does Not Cover: • Replacement parts or labor furnished by anyone REPAIR other than the Company, the selling dealer or a Service Company service technician When the Intelect® Mobile 2 or any accessories require •...
  • Page 59: Electromagnetic Compatibility (Emc) Tables

    APPENDIX ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES Guidance and manufacturer’s declaration – electromagnetic emissions The Intelect® Mobile 2 is intended for use in the electromagnetic environment specified below. The customer or the user of the Intelect® Mobile 2 should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
  • Page 60 APPENDIX ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED) Guidance and manufacturer’s declaration – electromagnetic immunity The Intelect® Mobile 2 is intended for use in the electromagnetic environment specified below. The customer or the user of the Intelect® Mobile 2 should assure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level...
  • Page 61 APPENDIX ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED) Guidance and manufacturer’s declaration – electromagnetic immunity The Intelect® Mobile 2 is intended for use in the electromagnetic environment specified below. The customer or the user of the Intelect® Mobile 2 should assure that it is used in such an electromagnetic environment. IMMUNITY test IEC 60601 TEST LEVEL Compliance level...
  • Page 62 APPENDIX ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED) Recommended separation distances between portable and mobile RF communications equipment and the Intelect® Mobile 2 The Intelect® Mobile 2 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Intelect® Mobile 2 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Intelect®...
  • Page 63 DJO France SAS Centre Européen de Frêt 3 rue de Bethar 64990 Mouguerre • France T: + 33 (0) 5 59 52 86 90 • F: + 33 (0) 5 59 52 86 91 DJOglobal.com DJO, LLC | 1430 Decision Street | Vista | CA 92081-8553 | U.S.A. www.DJOglobal.com...