Chattanooga Intelect TranSport 2783 User Manual

Chattanooga Intelect TranSport 2783 User Manual

Two channel transportable electrotherapy unit

Advertisement

Leading Therapy
User Manual
in a New Direction
Operation & Installation
Instructions for:
2783- Two Channel
Transportable
Electrotherapy Unit
© 2005 Encore Medical, L.P.

Advertisement

Table of Contents
loading

Summary of Contents for Chattanooga Intelect TranSport 2783

  • Page 1 Leading Therapy User Manual in a New Direction Operation & Installation Instructions for: 2783- Two Channel Transportable Electrotherapy Unit © 2005 Encore Medical, L.P.
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Intelect® Transport 2-Channel Electrotherapy FOREWORD ........1 CHARGING THE BATTERY PACK .
  • Page 3 TABLE OF CONTENTS Intelect® Transport 2-Channel Electrotherapy Restoring Factory Settings ......47 Viewing Unit Version Information ....48 TROUBLESHOOTING .
  • Page 4: Foreword

    This equipment is to be used only under the prescription and supervision of a licensed practitioner. ©2005 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial, or layout composition of this publication without express written consent from Chattanooga Group...
  • Page 5: Safety Precautions

    SAFETY PRECAUTIONS Intelect® Transport 2-Channel Electrotherapy PRECAUTIONARY DEFINITIONS The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definition of these symbols are as follows: Caution- Danger- Text with a “CAUTION”...
  • Page 6: Cautions

    DO NOT operate the Intelect Transport 2-Channel • The Intelect battery pack is designed for use only Electrotherapy unit when connected to any unit other than with Chattanooga Group Intelect Transport 2-Channel Chattanooga Group devices. Electrotherapy, Combo, Laser, and Ultrasound systems. •...
  • Page 7 SAFETY PRECAUTIONS Intelect® Transport 2-Channel Electrotherapy CAUTIONS continued • • The battery pack should be removed when storing the unit This equipment generates, uses and can radiate radio for extended periods of time. frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to •...
  • Page 8: Warnings

    SAFETY PRECAUTIONS Intelect® Transport 2-Channel Electrotherapy WARNINGS • These devices are restricted to sale by, or on the order of, • Before administering any treatment to a patient you should a physician or licensed practitioner. This device should be become acquainted with the operating procedures for used only under the continued supervision of a physician or each mode of treatment available, as well as the indications, licensed practitioner.
  • Page 9 SAFETY PRECAUTIONS Intelect® Transport 2-Channel Electrotherapy WARNINGS (continued) • Stimulation should not be applied over, or in proximity to, cancerous lesions. • Output current density is inversely related to electrode size. Improper application may result in patient injury. If any question arises as to the proper electrode size, consult a licensed practitioner prior to therapy session.
  • Page 10: Dangers

    LCD beginning with a 2 or 3, immediately stop all use of the pack. Doing so can increase the individual cell unit and contact the dealer or Chattanooga Group for service. temperature and cause cell rupture or leakage. Errors and Warnings in these categories indicate an internal •...
  • Page 11 SAFETY PRECAUTIONS Intelect® Transport 2-Channel Electrotherapy DANGERS continued • Dispose of NiMH batteries according to national, state, and local codes and regulations.
  • Page 12: Indications, Contraindications, And Adverse Effects

    SAFETY PRECAUTIONS Intelect® Transport 2-Channel Electrotherapy INDICATIONS, CONTRAINDICATIONS, AND ADVERSE EFFECTS • Safety has not been established for the use of therapeutic Indications for Russian, High Voltage Pulsed Current (HVPC), electrical stimulation during pregnancy. IFC, and Premodulated Waveforms • • Relaxation of muscle spasms Powered muscle stimulators should not be used on patients with cardiac demand pacemakers.
  • Page 13: Adverse Effects

    SAFETY PRECAUTIONS Intelect® Transport 2-Channel Electrotherapy INDICATIONS, CONTRAINDICATIONS, AND ADVERSE EFFECTS (continued) Adverse Effects • Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators. • Potential adverse effects with TENS are skin irritation and electrode burns.
  • Page 14: Overview

    Intelect® Transport 2-Channel Electrotherapy The Intelect Transport 2-Channel Electrotherapy unit, designed and manufactured by Chattanooga Group, offers a new dimension in portable electrotherapy made possible by advanced software design and digital signal processing. The result is a unit with extraordinary versatility based on simplicity of operation.
  • Page 15 OVERVIEW Intelect® Transport 2-Channel Electrotherapy Burst Frequency (Freq.) - This is the number of bursts per second (bps). The available burst frequencies on the Intelect Transport 2- Channel Electrotherapy unit are 20 to 100 bps. This parameter is unique to the Russian waveform. Carrier Frequency (Freq.) - Associated with the Interferential and Russian waveforms, Carrier Frequency is the frequency of the un- modulated medium frequency current.
  • Page 16 OVERVIEW Intelect® Transport 2-Channel Electrotherapy Intensity - Intensity is the output of electrotherapy distributed by the unit to the patient. Depending on the waveform, intensity is measured in milliamps (mA), volts (V), and microamps (μA). Leadwires - The leadwires consist of the main plugs that are connected to the unit, and 4 leads (2 black and 2 red) that connect to electrodes.
  • Page 17 OVERVIEW Intelect® Transport 2-Channel Electrotherapy Waveforms - Waveforms are current or voltage that is varied by time and are the geometrical descriptions of a DC, AC, or pulsed DC/AC current. Current waveforms are described as either monophasic or biphasic. A biphasic wave is further described as either symmetrical or asymmetrical and as balanced or unbalanced.
  • Page 18: Nomenclature

    NOMENCLATURE Intelect® Transport 2-Channel Electrotherapy LCD Intensity/ Power Contrast Dial On/Off Arrow Clinical Resources INTENSITY TIME Parameter Display/Enter Channel 1 Channel 2 Back Lead Wire Lead Wire Connection Connection START STOP PAUSE Down Accessory Arrow Panel...
  • Page 19 NOMENCLATURE Intelect® Transport 2-Channel Electrotherapy Power On/Off The Power On/Off button controls the flow of electricity to the unit. NOTE: Make certain there are no electrodes on the patient when turning the unit on or off. The LCD (Liquid Crystal Display) allows the user to view and monitor the information displayed before, during, and after therapy. Clinical Resources Select this button to access the following functions: •...
  • Page 20 NOMENCLATURE Intelect® Transport 2-Channel Electrotherapy Accessory Panel The Accessory Panel serves as a port of connection for the electrodes. Channel 1 Lead Wire Connection This port serves as the connection point between the unit and the Channel 1 Lead Wire. Channel 2 Lead Wire Connection This port serves as the connection point between the unit and the Channel 2 Lead Wire.
  • Page 21 NOMENCLATURE Intelect® Transport 2-Channel Electrotherapy Charge Indicator This symbol displays when the unit is connected to mains power and the battery pack is charging. NOTE: During battery operation, if the unit is left on, but is not active for more than five minutes, it will power off to conserve battery power.
  • Page 22: Specifications

    SPECIFICATIONS Intelect® Transport 2-Channel Electrotherapy UNIT SPECIFICATIONS Dimensions Width ..........11.3 in (28.8 cm) Height .
  • Page 23: Waveform Specifications

    SPECIFICATIONS Intelect® Transport 2-Channel Electrotherapy WAVEFORM SPECIFICATIONS Premodulated IFC Premodulated current is a medium frequency waveform. Current IFC (Interferential) current is a medium frequency waveform. comes out of one channel (two electrodes). The current intensity Current is distributed through two channels (four electrodes). is modulated: it increases and decreases at a regular frequency The currents cross each other in the body at the area requiring (the Amplitude Modulation Frequency).
  • Page 24 SPECIFICATIONS Intelect® Transport 2-Channel Electrotherapy WAVEFORM SPECIFICATIONS (continued) High Voltage Pulsed Current (HVPC) Russian The High Voltage Pulsed Current (HVPC) has a very brief pulse Russian Current is a sinusoidal waveform, delivered in bursts or duration characterized by two distinct peaks delivered at series of pulses.
  • Page 25: Setup

    SETUP Intelect® Transport 2-Channel Electrotherapy MOUNTING THE UNIT ON THE WALL The Intelect® Transport 2-Channel Electrotherapy unit can be operated while the unit is resting on a flat surface or mounted on a wall. To mount the unit on a wall, do the following: Remove the repositional base from the back of the unit.
  • Page 26 SETUP Intelect® Transport 2-Channel Electrotherapy MOUNTING THE UNIT ON THE WALL continued Using an appropriate size drill bit, drill four holes you marked in the previous step. Press 4 appropriately sized drywall anchors into the wall so that the drywall anchor is flush with the wall.
  • Page 27 SETUP Intelect® Transport 2-Channel Electrotherapy MOUNTING THE UNIT ON THE WALL continued Screw four #8 pan head sheet metal screws (1 inch or 2.54 cm) into the wall anchors. Make sure you leave 1/4 of an inch (0.635 cm ) between the wall and the head of the screw.
  • Page 28 SETUP Intelect® Transport 2-Channel Electrotherapy MOUNTING THE UNIT ON THE WALL continued Line up the screw heads with the holes on the repositional base, and slide the unit down slightly until the screw heads are securely fastened to the repositional base.
  • Page 29: Installing The Battery Pack

    SETUP Intelect® Transport 2-Channel Electrotherapy INSTALLING THE BATTERY PACK The Intelect® Transport 2-Channel Electrotherapy unit accommodates both AC mains power and an optional battery pack. The pack contains 20 Nickel Metal Hydride (NiMH) drycell batteries. To install the battery pack in the unit, do the following: Locate the battery access door at the bottom of the unit and loosen the screw with a flat head screwdriver.
  • Page 30 SETUP Intelect® Transport 2-Channel Electrotherapy INSTALLING THE BATTERY PACK continued Connect the battery pack cable to the unit’s battery connector in the bottom of the battery recess. Put the battery pack into the unit, making sure to orient it as shown.
  • Page 31: Charging The Battery Pack

    SETUP Intelect® Transport 2-Channel Electrotherapy INSTALLING THE BATTERY PACK continued Replace the battery access door and re-tighten the screw using the screwdriver. Reverse the steps in this section in order to remove the battery pack. CHARGING THE BATTERY PACK The battery pack is automatically charged by the unit whenever there is mains power connected. Charging may be interrupted during operation of the unit by the control circuitry to limit total power consumption.
  • Page 32: Patient Preparation

    PATIENT PREPARATION Intelect® Transport 2-Channel Electrotherapy ELECTROTHERAPY PATIENT PREPARATION Electrode Placement Use the following guidelines when preparing patients for electrotherapy: • Examine the skin for any wounds and clean the skin. • Keep electrodes separated during treatment. Electrodes in contact •...
  • Page 33: Dura-Stick Electrodes

    ELECTROTHERAPY PATIENT PREPARATION (continued) DURA-STICK Electrodes Reusable Carbon Electrodes Chattanooga Group Dura-Stick Electrodes are a self adhesive, If used for delivery of electrotherapy, the Carbon Electrodes single patient, one time use disposable product designed must be used with a conductive medium such as Conductor™...
  • Page 34: Dura-Stick Electrode Instructions

    PATIENT PREPARATION Intelect® Transport 2-Channel Electrotherapy ELECTROTHERAPY PATIENT PREPARATION (continued) DURA-STICK Electrode Instructions Connecting Lead Wires Securing Electrodes Insert the lead with the Red (+) electrode NOTE: Use of conductive medium or Remove the Dura-Stick Electrodes from connector into one Dura-Stick Electrode. sponges is not required or recommended.
  • Page 35: Reusable Carbon Electrodes

    PATIENT PREPARATION Intelect® Transport 2-Channel Electrotherapy ELECTROTHERAPY PATIENT PREPARATION (continued) Reusable Carbon Electrodes Conductive Medium Securing Electrodes Connecting Lead Wires Insert the lead with the Red (+) electrode Liberally apply Conductor™ Transmission Use the Nylatex Wrap to secure each ® connector into Red electrode.
  • Page 36: Operation

    OPERATION Intelect® Transport 2-Channel Electrotherapy STARTING, STOPPING, AND INTERRUPTING THERAPY The Operator Interface consists of buttons with a liquid crystal display (LCD). The operator is able to view parameter options on the display and make selections by pressing the buttons on the control panel. The LCD will provide continuous information during the treatments concerning power and elapsed time.
  • Page 37 OPERATION Intelect® Transport 2-Channel Electrotherapy STARTING, STOPPING, AND INTERRUPTING THERAPY continued Depending on the type of waveform you intend to use and the number of patients you intend to treat, insert the Lead Wire into Channel 1, Channel 2, or both Lead Wire Connections on the Accessory Panel.
  • Page 38 OPERATION Intelect® Transport 2-Channel Electrotherapy STARTING, STOPPING, AND INTERRUPTING THERAPY continued Use the Up and Down Arrow buttons to highlight the appropriate waveform. Press the Enter button. The Parameter screen displays. On the parameter screen, use the Up and Down Arrow buttons to highlight the parameter you want to change and adjust it accordingly, making sure to press the Enter button after each adjustment.
  • Page 39 OPERATION Intelect® Transport 2-Channel Electrotherapy STARTING, STOPPING, AND INTERRUPTING THERAPY continued To begin electrotherapy, press the START button. The timer counts down, the output power ramps up, and "Running" displays below the timer. NOTE: When the therapy time has expired, the unit beeps three times. During therapy, you can press the TIME button to raise or lower treatment time (in one minute increments) using the up and down arrows.
  • Page 40: Creating A User Protocol

    OPERATION Intelect® Transport 2-Channel Electrotherapy CREATING A USER PROTOCOL This is a library you create. You may store up to 15 protocols in the User Protocol Library. To create User Protocols, do the following: On the home screen, press the Enter button. NOTE: User protocols can be used on either channel.
  • Page 41 OPERATION Intelect® Transport 2-Channel Electrotherapy CREATING A USER PROTOCOL continued Use the Up Arrow and Down Arrow buttons to highlight any unused user protocol. If you select the Save as Default protocol, this will become the protocol displayed when the waveform is selected on the Waveform screen. Press the Enter button to accept the highlighted selection and save your custom protocol.
  • Page 42: Using Protocols

    OPERATION Intelect® Transport 2-Channel Electrotherapy USING PROTOCOLS To utilize a user protocol, do the following: On the home screen, press the Clinical Resources button. The Clinical Library screen displays. Using the Up Arrow and Down Arrow buttons, highlight the Retrieve User Protocol option.
  • Page 43 OPERATION Intelect® Transport 2-Channel Electrotherapy USING PROTOCOLS continued To begin therapy, perform all the procedures outlined in the section entitled "Patient Preparation" page 29. Then continue with step 11 of the section entitled "Starting, Stopping, and Interrupting Therapy" page...
  • Page 44: System Utilities

    OPERATION Intelect® Transport 2-Channel Electrotherapy SYSTEM UTILITIES Audible Tones Audible tones will be heard in the following conditions: • Any button is pressed. • The rechargeable battery’s power is low (in which case the Low Battery icon will display). • Any error message is displayed. •...
  • Page 45: Changing Default Protocols

    OPERATION Intelect® Transport 2-Channel Electrotherapy SYSTEM UTILITIES continued Using the Up Arrow and Down Arrow buttons, highlight the appropriate parameter and make the necessary changes, making sure to press the Enter button after each adjustment. Changing Default Protocols To change the power up presets of the waveforms, do the following: On the home screen, press the Enter button.
  • Page 46 OPERATION Intelect® Transport 2-Channel Electrotherapy SYSTEM UTILITIES continued Use the Up and Down Arrow buttons to highlight the appropriate waveform. Press the Enter button. The Parameter screen displays. On the parameter screen, use the Up and Down Arrow buttons to highlight the parameter you want to change and adjust it accordingly.
  • Page 47: Brightening Or Dimming The Lcd

    OPERATION Intelect® Transport 2-Channel Electrotherapy SYSTEM UTILITIES continued Press the Enter button to accept the highlighted selection. The Default Protocol Confirmation window displays. Press any key to confirm the settings. You are returned to the Clinical Resources menu. Brightening or Dimming the LCD To brighten or dim the LCD, turn the contrast control dial until the display contrast is optimal.
  • Page 48: Restoring Factory Protocols

    OPERATION Intelect® Transport 2-Channel Electrotherapy SYSTEM UTILITIES continued Restoring Factory Protocols If necessary, you can choose to restore the unit’s original (default) waveform parameters when it was shipped to you. NOTE: This procedure will erase all user-defined protocols. To restore the unit’s original waveform parameters, do the following: Press the Clinical Resources button.
  • Page 49: Changing Languages

    OPERATION Intelect® Transport 2-Channel Electrotherapy SYSTEM UTILITIES continued Changing Languages You may change the language displayed by the Intelect® Transport 2-Channel Electrotherapy unit to either English or Spanish. To change the language displayed on the LCD, do the following: Press the Clinical Resources button. The Clinical Resources screen displays.
  • Page 50: Restoring Factory Settings

    OPERATION Intelect® Transport 2-Channel Electrotherapy SYSTEM UTILITIES continued Restoring Factory Settings To restore the original language on the unit, do the following: On the main window, press the Clinical Resources button. The Clinical Library screen displays. Press the Up Arrow or Down Arrow buttons to highlight the Restore Factory Settings option.
  • Page 51: Viewing Unit Version Information

    OPERATION Intelect® Transport 2-Channel Electrotherapy SYSTEM UTILITIES continued Viewing Unit Version Information Use this utility to determine the unit’s software version. To do this, do the following: Press the Clinical Resources button. The Clinical Library screen displays. Use the Up Arrow and Down Arrow buttons to highlight the View Unit Information option.
  • Page 52: Troubleshooting

    TROUBLESHOOTING Intelect® Transport 2-Channel Electrotherapy ERROR CODES The Intelect Transport 2-Channel Electrotherapy unit incorporates error messages and warnings to inform the user of problems or potential problems with the unit, modality, or accessories. These are numbered so the user can possibly correct the problem without the aid of service personnel.
  • Page 53 In the event that an Error message or Warning appears on the LCD beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer or Chattanooga Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested by Chattanooga Group or a Field Service Technician certified by Chattanooga Group before any further operation or use of the system.
  • Page 54: Accessories

    ACCESSORIES Intelect® Transport 2-Channel Electrotherapy Standard Accessories Optional Accessories Ref. Description Ref. Description 27478 NiMH Battery Pack 27376 Electrotherapy Accessory Kit- Includes the following: 27467 Intelect Transport 2 Channel Electrotherapy Carrying Bag 27312 Channel 1 Lead Wire 27313 Channel 2 Lead Wire 10648 Nylatex®...
  • Page 55: Maintenance

    FACTORY SERVICE When the Intelect Transport 2-Channel Electrotherapy unit requires Cleaning the Unit and the Accessories factory service, contact the selling dealer or Chattanooga Group With the unit disconnected from the power source, clean the Service Department. unit with a clean, lint free cloth moistened with water and mild WARRANTY REPAIR/OUT OF WARRANTY REPAIR antibacterial soap.
  • Page 56: Warranty

    Intelect® Transport 2-Channel Electrotherapy Chattanooga Group ("Company") warrants that the Intelect Transport 2-Channel Electrotherapy unit ("Product") is free of defects in material and workmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If this Product fails to function during the two year warranty period due to a defect in material or workmanship, Company or the selling dealer will repair or replace this Product without charge within a period of thirty (30) days from the date on which the Product is returned to the Company or the dealer.
  • Page 57 Leading Therapy in a New Direction ISO 13485 Certified 4717 Adams Road P.O. Box 489 Hixson, TN 37343 U.S.A. (800) 592-7329 - U.S.A. (800) 361-6661 - Canada (800) 242-8329 FAX www.chattgroup.com 27936A © 2005 Encore Medical, L.P.