MAINTENANCE
CARING FOR THE THERAPY SYSTEM
Cleaning the Therapy System
With the system disconnected from the power source, clean the
system with a clean, lint free cloth moistened with water and mild
antibacterial soap. If a more sterile cleaning is needed, use a cloth
moistened with an antimicrobial cleaner.
Do not submerse the system in liquids. Should the unit
accidentally become submersed, contact the dealer or DJO Service
Department immediately. Do not attempt to use a system that
has been wet inside until inspected and tested by
a Service Technician Certified by DJO.
Do not allow liquids to enter the ventilation holes in the optional
modules. This could permanently damage the modules.
Cleaning Electrode Sponges
Thoroughly clean sponges after each use with medical grade alcohol.
Cleaning the Lens
Clean the Therapy System Screen Lens using NOVUS® Polish System.
Contact Novus at: www.novuspolish.com
CALIBRATION REQUIREMENTS
Calibrating Ultrasound Applicators
Annual factory calibration is required for all Ultrasound Applicators.
Only the Applicators should be sent to the factory or a Field
Technician certified by DJO for this procedure.
NOVUS is the Registered Trademark of NOVUS Inc.
SERVICE
Should the unit require service, warranty, or repair, please contact
the selling dealer or your local DJO customer service.