OUTLOOK " 200 wir Dosescan
OPERATOR'S MANUAL
OPTIONS
The Outlook™ 200 offers other features which allow the pump to be customized to each
patient's particular needs.
δν
NOTE: These features may be found in either the Main or
"
Alternate Menus, or may have been disabled and
therefore unavailable.
PANEL LOCK OUT SWITCH
This option allows the keypad to be deactivated. While Panel Lock Out is activated, no keys
are functional.
1. While the pump is infusing, press the Panel Lock Out Switch on the back of the pump.
2.
To unlock the keypad, press the Panel Lock Out Switch again.
CLEAR TOTAL INFUSED
1.
Press MENU.
2. Use the CURSOR to highlight "Clear Total Infused," then
The PS
Total Infused
press ENTER. The Total Infused value will be displayed.
이
I 5 this OK?
The Information Screen will display a message asking
if the Total infused should be cleared.
YES
NO
3.
Use the CURSOR to highlight "YES," then press ENTER.
Cursor <--> & Press ENTER]
3
Note: The Total infused may be cleared while the pump is
infusing.
OCCLUSION LIMIT
This feature allows the user to change the occlusion limit to 75, 100, 200, 300, 400, or 500
mmHg. The Occlusion Limit is a threshold value.
CAUTION: The higher the pressure limit, the less sensitive
the pump is to changes in fluid resistance which may be
caused by positional IVs or infiltrations. High pressure
limits can increase the severity of an infiltration without an
alarm condition.
1.
Press HOLD to stop the infusion.
2.
Press MENU.
3.
Use the CURSOR to highlight "Occlusion Limit," then press ENTER.
50
950952 Rev A (9/03)