A,
Chapter 10: Options
4.
Use the DATA keys to change the occlusion limit value.
5.
Press ENTER.
EXAMPLE
If the Occlusion Limit is set to 300mmHg, the pump will infuse at the lowest pressure pos-
sible but will alarm if the pressure required to infuse the IV fluid rises above the
300mmHg value.
NOTE: A gravity infusion of D5W with the IV container 3
A
feet above the IV site is approximately 75 mmHg.
NoTE: The Occlusion Limit will automatically change to
400 mmHg to accommodate fluid resistance at rates
"~
greater than 400 mi/hr.
OCCLUSION WARNING
The occlusion warning feature gives an audible warning (3 beeps) and a visual indication
in the Volume To Be Delivered Display to let the user know an occlusion is occurring
downstream from the pump. The warning occurs 6 times at 5 second intervals before the
pump alarms "Occlusion." If the occlusion is cleared before alarming, the warning tones
will stop and the infusion will continue.This feature is designed to alert the user of down-
stream resistance before an alarm occurs which stops the infusion.
During an occlusion warning, inspect the downstream tubing for kinks or closed clamps.
Remove the kinks or open the clamps and the warning tones will stop.
NoTE: The Occlusion Warning option must be turned on in
the Biomed Menu by an authorized person.
SET TIME AND VOLUME
This option calculates the rate of the infusion when the infusion time and volume are entered.
1.
2.
3.
4.
Press MENU.
Use the CURSOR to highlight "Set Time and Volume,"
then press ENTER.
一
Use the CURSOR to highlight Time= "hr" or "min"
Rate = 75.0 mMhr
Press DATA keys to enter the correct hours and/or min-
Time = 01 hr 40 min
utes,
Volume = [Emi
Use the CURSOR to highlight Volume= field.
ACCEPT
Use the DATA keys to enter the volume.
Use the CURSOR to highlight "ACCEPT", then press ENTER.
Set Time and Volume
950952 Rev A (9/03)
51