Warnings
ARNINGS
The following warnings should be followed in order to avoid patient/user injury.
WARNING
'WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
This Operator's Manual contains detailed instructions and warnings on
the use of the Outlook 200. Please read it completely prior to using this
device. This manual is intended to reinforce the teaching given to the
user by a trained health care professional or an authorized B. Braun
representative.
B. Braun will assume no responsibility for incidents which may occur
if the product is not used in accordance with product labeling.
The Outlook 200 has been designed to stop fluid flow to the patient
under alarm conditions. Monitoring of the patient and infusion status is
necessary to ensure the infusion is being delivered as anticipated. Prior
to starting an infusion, verify no drops are falling in the drip chamber
and the programmed information is correct. If drops are falling and the
infusion has not begun, the set may be improperly loaded and further
use may result in inaccurate infusion.
Do not connect the IV administration set to the patient while priming,
as a bolus or over infusion may occur.
See EPIDURAL in DOSE Mode in Chapter 6: DOSE Mode for spe-
cific warnings and cautions related to epidural medication infusions.
To avoid mechanical or electronic damage, do not steam autoclave or
immerse the pump in any fluids or cleaning solutions. Always discon-
nect the electrical power cord from the outlet before cleaning to prevent
electrical shock.
Trained Biomedical professionals or B. Braun representatives must per-
form a full set-up of the pump before use in a clinical setting to ensure
proper programming and function of device.
Do not use in the presence of flammable anesthetics, as a possible
explosion hazard exists.
Avoid exposure to powerful sources of electromagnetic interference
and strong magnetic fields such as Magnetic Resonance Imaging
(MRI). Patient injury and equipment damage may result.
Device contains Class 2 laser product. Laser radiation present. Do not
stare into beam.
950952 Rev A (9/03)