Chapter 9: Profile Mode
ENTER. A graph of the infusion will be displayed.
9.
The TEXT field in the upper right corner of the Information Screen is highlighted; press
ENTER. The text display will return.
10. If the pump is turned off while in PROFILE Mode, the Information Display will indicate
"PROFILE Operation was Interrupted" when the pump is turned back on. Use the CURSOR to
highlight one of the following, then press ENTER.
+
Resume current PROFILE
*
Restart PROFILE
*
Exit PROFILE Mode
REVIEW/CHANGE
PROFILE
1.
Press MENU.
2.
Use the CURSOR to highlight "Review/Change PROFILE," then press ENTER.
3. Use the CURSOR to highlight the Volume field.
^
NOTE: The PROFILE Volume cannot be changed after
"
PROFILE Mode has been started.
Use the DATA keys to enter the new volume.
Use the CURSOR to highlight the Time field.
Use the DATA keys to enter the new time.
Use the CURSOR to highlight "ACCEPT PROFILE," then press ENTER.
RESTART PROFILE
Note: Choosing Restart PROFILE will cance] the present
*
infusion and the pump will begin at the beginning of the
Profile run.
Press HOLD.
Press MENU.
Use the CURSOR to highlight "Restart PROFILE," then press ENTER.
A screen will appear with the message: The "Restart PROFILE" option will cancel the present
run. Is this OK? Use the CURSOR to highlight "YES," then press ENTER.
950952 Rev A (9/03)
47