OUTLOOK™ 200 with Dosescan" OPERATOR'S MANUAL
UNLOADING THE TUBING
Press HOLD to stop the infusion.
Close the roller clamp on the tubing.
Open the pump door.
Grasp the Free Flow Protection Clip and the tubing on the left hand side of the pump.
Gently pull the Free Flow Protection Clip from the Free Flow Protection Clip Receptacle
while also pulling the tubing/cassette from the pump. Removing the Free Flow Protection
Clip in this manner will automatically stop the flow of fluid in the tubing.
6.
If gravity infusion is desired, close the roller clamp, open the Free Flow Protection Clip,
and adjust the flow with the roller clamp.
TURNING
OFF THE PUMP
The Outlook™ 200 must be stopped before turning the device off.
1.
Press HOLD to stop current infusion.
2.
Press POWER.
14
950952 Rev A (9/03)