Chapter 6: DOSE Mode
TO CHANGE UNITS ON THE DATA ENTRY SCREEN
Critical medications require units for calculation of doses, for example mcg/mg/min.
Three categories of units are Dosage, Concentration (Conc), and Body. Each of these can
be independently selected and changed. Match the dosage and concentration as ordered by
the physician.
1.
Use the CURSOR to highlight "SETUP" then
DOSAGE
CONC
BODY
press ENTER.
.
2. Use the left and right CURSOR keys to
megikgimin
Omg
kg
highlight the parameter to be changed.
3.
Use the up and down CURSOR keys to
change the parameter.
4. Press ENTER.
Press ENTER when done
; change
<- -> move
CHANGING DOSE MODE RATE
The rate may be titrated while the pump is infusing.
1.
Use the DATA keys to change the rate. The dose will be recalculated depending on the new
rate entered.
2.
Press ENTER or PRIMARY RUN to confirm the new rate and dose and begin the infusion.
NOTE: İf ENTER or PRİMARY RUN is not pressed within 4
seconds of entering a new rate and dose, a series of error
beeps occurs and the rate/dose change is cancelled. The
pump then displays the message "Rate change aborted,"
and continues to infuse at the old rate. An intermittent alarm
continues to sound to notify the user a rate change was
attempted and cancelied. Pressing any key will silence the
alarm.
CHANGING DOSE MODE DOSE
The dose can be titrated while the pump is infusing.
1.
Use the CURSOR to highlight the DOSE field in the
・
.
D
i
DOSE
upper right corner of the Information Screen, then press
opamine
.
ENTER.
5.00 mcg/kg/min
2. On the DATA ENTRY SCREEN, use the CURSOR to
|TmeLeft=
13hr
28min
highlight the Dose field.
Total Infused =
136.9 ml
3.
Use the DATA keys to change the dose, then press
Occlusion Limit=
300mmHg
ENTER.
4.
Use the CURSOR to highlight "ACCEPT," then press ENTER. The rate will be recalculated
depending on the new dose entered.
950952 Rev A (9/03)
33