OUTLOOK " 200 win DosEScAN " OPERATOR'S MANUAL
ALARM
MESSAGE
CAUSES
CORRECTION
System Error
Error
This alarm is generated
when there has been an
internal pump malfunction.
Turn the pump off and
return to Biomedical Pro-
fessional for service
Set Primary
Rate
Flashing 0.0
This alarm is generated
when the user tries to start
a primary channel infusion
and has not set a primary
rate.
Set a Primary Rate.
Set Piggyback
Rate
Flashing 0.0
This alarm is generated
when the user tries to start
a piggyback channel infu-
sion and has not set a pig-
gyback rate.
Set a Piggyback Rate.
Set Primary
Volume
Flashing 0.0
This alarm is generated
when the user tries to start
a primary channel infusion
and has not set a primary
VTBD.
Set a Primary VTBD.
Set Piggyback
Volume
Flashing 0.0
This alarm is generated
when the user tries to start
a piggyback channel infu-
sion and has not set a pig-
gyback VTBD.
Set a Piggyback VBTD.
62
950952 Rev A (9/03)