OUTLOOK ™ 200 wir Dosescan " OPERATOR'S MANUAL
NOTE: To prevent incorrect delivery, the programmed
piggyback volume to be delivered must be equal to the
amount of fluid in the piggyback container. Piggyback
Mode is not intended for the infusion of fluids requiring
flushing before or after administration.
6.
Press PIGGYBACK RUN to start the infusion.
7.
When the VTBD of the Piggyback channel goes to zero, the pump will change back to the
Primary channel RATE and VTBD. Make sure the primary tubing is not clamped.
CAUTION: Do not attempt to infuse both primary and
piggyback fluids simultaneously using this method;
sympathetic flow may result. Use primary sets with check
valves.
NOTE: The pump must be stopped to activate Piggyback
Mode.
NOTE: ff the piggyback container is not empty, the
remaining piggyback fluid will be delivered at the Primary
channel rate after the pump changes back to the Primary
channel.
20
950952 Rev A (9/03)