OUTLOOK™ 200 wir Dosescan " OPERATOR'S MANUAL
STANDARD
MODE
STANDARD Mode allows for infusion of fluids when rate and volume to be
delivered data is entered. At the end of the infusion, the pump automatically
switches to Keep Vein Open in order to maintain IV patency. In this mode,
infusion
secondary fluid, such as an antibiotic, may be administered using a
PIGGYBACK method of delivery. Piggyback infusions are allowed only in
STANDARD Mode.
1.
Press PRIMARY CHANNEL
INDICATOR
©.
ρω
μη
2.
Use the DATA keys to enter the rate and volume to be
[rar
PRIMARY
delivered.
©
125.0!
1000.0
3. Press PRIMARY RUN to start the infusion. The green
[Soh
PIGEN RACK
Run Indicators will light up.
4.
If desired, press the Panel Lock Out Switch on the
back of the pump to make the keypanel tamper proof.
WARNING: After pressing PRIMARY RUN, make sure drops
are falling in the drip chamber. If no drops are falling, make
sure the roller clamp is open. If the roller clamp is open
and still no drops are falling, remove and throw away the
set as it may be defective.
The Time Left, Total Infused, and Occlusion Limit are shown on the Information Screen.
Monitor the infusion according to hospital policy. After the infusion is complete, the pump
will alarm and go into KVO.
CHANGING THE RATE
With Pump infusing
1.
Press the DATA keys to change the rate.
2.
Press ENTER Or PRIMARY RUN to confirm new rate.
NOTE: If ENTER or PRIMARY RUN is not pressed within 4
seconds of entering a new rate, a series of error beeps will
occur and the rate change will be cancelled. The pump will
then display the message "Rate change aborted," and will
continue to infuse at the old rate. An intermittent alarm will
continue to sound to notify the user that a rate change has
been attempted and cancelled. Pressing any key will
silence the alarm.
18
950952 Rev À (9/03)