POLE CLAMP OPERATION
Press down on the pole clamp lever.
Position the pump on the pole (make sure the pole is up against the slot of the pump).
Release the pole clamp lever.
Lock by rotating the pole clamp lever 1/4 turn toward the back of the pump.
To remove the pump from the pole, rotate the pole clamp lever 1/4 turn toward the front of
the pump to unlock it, then press down on the pole clamp lever.
TURNING ON THE PUMP
1.
Plug the pump into a hospital-grade electrical outlet or the AC Receptacle on the back of
another Outlook™ 200 pump.
2.
Press POWER.
NOTE: The Outlook 200 will perform a self test which ensures
"
proper system operation and alarm reporting. Within this
process, the AC Power Indicator will light up, the pump will
beep, and the Information Screen will light up. A successful
self-test allows safe operation of the device.
DAISY CHAINING
Daisy Chaining allows up to 5 Outlook 200 pumps to be plugged into each other with only
one pump plugged into a wall outlet. On the back of each Outlook 200 is the access door to an
electrical outlet. Only another Outlook 200 pump should be plugged into that outlet.
ATTENTION: The Daisy Chain outlet is for use with Outlook
pumps for the purpose of daisy chaining only. Do not daisy
chain more than 5 pumps, as the total leakage current of a
system containing more than 5 pumps may exceed 300 uA.
BATTERY OPERATION
The Outlook 200 has been equipped with an internal, rechargeable battery which will power
the pump automatically if the plug is disconnected during use. Battery power is provided as a
back-up system only, and the length of the battery life is affected by the rate of delivery.
Whenever possible, the pump should remain plugged into an electrical outlet. The Battery
Indicator will be lit when the pump is operating on its battery.
The pump must remain plugged into a hospital-grade electrical outlet for at least 24 hours to
fuily recharge the battery from a totally discharged condition.