OUTLOOK." 200 wrm Dosrscax " OPERATOR'S MANUAL
DoseScan™
DoseScan™ activates either the bar code scanner located in the handle of the
|!
pump or an optional hand held scanner. Information contained on the bar
Doe, John O.
coded drug label will be used to configure the infusion in STANDARD,
EA INI
Piggyback, or DOSE Mode. The facility has the option to choose if an
Normal Saline
Authorized User ID and Patient ID are required for data entry. The pump only
ΙΙ
prompts the user to scan in required information.
Use of controls or adjustments or performance of
procedures other than those specified herein may result in
hazardous radiation exposure.
FOR STANDARD
MODE & DOSE MODE INFUSIONS
1.
Press MENU.
2.
Use the CURSOR to highlight "DoseScan™," then press ENTER.
Note: If the hand held scanner is used, press the trigger
on the scanner after each line is successfully scanned.
The scanner will only remain active for 30 seconds.
If the
scanner turns off, the user will need to reactivate the
scanner.
%
NOTE: SCANNING SEQUENCE:
à
1) authorized user identification,
2) patient identification, and
3) infusion data information from the infusion container.
NOTE: An audible tone will confirm the data has been
accepted with each pass in front of the DoseScan scanner.
Scan Authorized User identification.
4.
The Information Display will ask if this is a new patient. Use the CURSOR to highlight
either "YES" or "NO", then press ENTER.
Scan Patient identification.
Scan Patient identification information on the medication label.
Scan the IV medication data on the medication label. The infusion data will be displayed
on the Information Display.
Review the data, highlight "ACCEPT", then press ENTER.
9.
Press RUN to start the infusion.
26