9.
Use the CURSOR to highlight the GRAPH field in the upper right corner of the Information
Screen, then press ENTER. A graph of the infusion will be displayed.
NOTE: Each period is represented by a vertical bar which
fills as the fluid is infused. The height of the bar represents
the rate of infusion in that particular period. The period
presently being infused is displayed in the upper left
corner. If the vertical bar appears to be missing, the rate in
that period is much less compared to the other periods.
10. Use the CURSOR to highlight the TEXT field in the upper right corner of the Information
Screen, then press ENTER. The text display will return.
11. If the pump is turned off while in PROGRAM Mode, the Information Display will indi-
cate "PROGRAM Operation was Interrupted" when the pump is turned back on. Use the CUR-
SOR to highlight one of the following, then press ENTER.
*
Resume current PROGRAM
*
Restart PROGRAM
-
Exit PROGRAM Mode
CLEARING PERIODS
The periods in PROGRAM Mode may be cleared.
1.
Use the CURSOR to highlight ZERO at the end of the period line to be cleared then press
ENTER.
2.
Repeat step 1 until all periods are cleared.
REVIEW/CHANGE PROGRAM
Note: Program Volumes cannot be changed after
PROGRAM
Mode has been started.
Press MENU.
Use the CURSOR to highlight "Review/Change PROGRAM," then press ENTER.
Use the CURSOR to highlight the Rate field of the period to be changed.
Use the DATA keys to enter the new rate.
Use the CURSOR to highlight "ACCEPT," then press ENTER.