OUTLOOK™ 200 wirk DoseScan™" OPERATOR'S MANUAL
DOSE
MODE
DOSE Mode calculates the concentration, rate, or dose information necessary
with many critical drug infusions. This data may include infusion specific
parameters such as patient weight, drug dosage, and drug concentrations.
Titration of the medication automatically leads to recalculation of the dosing
parameter. The pump contains a library of commonly dosed IV drugs which
may be customized into area specific categories and concentrations.
DOSE Mode may be used with or without a drug name label. Displaying the drug name, rate,
and dose allows the user to view valuable information. As a safety precaution, DOSE Mode
cannot be used with any other operating modes. The user must accept all data entries prior to
beginning an infusion.
Note: DOSE Mode is a calculation and labeling feature
only. Physician orders will always govern the appropriate
patient dosage.
Note: DOSE Mode may be placed in either the Main or
* Alternate Menus.
1.
Press HOLD.
2.
Press MENU.
3.
Use the CURSOR to highlight "DOSE Mode," then press ENTER.
One of two screens will appear, depending on whether or not DOSE Mode was used
previously.
2.
See "DATA ENTRY SCREEN."
TO CHANGE TO A DIFFERENT DRUG
1.
Use the CURSOR to highlight "Change," then press ENTER.
2. Use the CURSOR to highlight the Drug Cat-
DRUG LISTS
egory desired, then press ENTER.
.
uE
Neo & Pedi
ed
Surg
Oncology
Anesthesia
Epidural
CLEAR DRUG|
[EXIT
30
950952 Rev À (9/03)