OUTLOOK ™ 200 witn DoseScan" OPERATOR'S MANUAL
PROGRAM
MODE
This mode is for infusions requiring a non-standard delivery pattern.
PROGRAM Mode allows up to 9 periods to be entered. Each period contains | Seventie! Periods
its own rate and volume. The pump automatically gives each programmed
period, one after the other.
Note: PROGRAM Mode may only be entered when the
pump is stopped in STANDARD Mode.
を 。
NOTE: PROGRAM Mode may be found in either the Main
or Alternate Menus, or may have been disabled and
therefore unavailable.
1.
Press MENU.
2.
Use the CURSOR to highlight "PROGRAM Mode," then press ENTER.
3.
Use the DATA keys to enter the rate and volume for the
Pd.
Rate
Volume
first period.
41
500
1000 ...
4.
Use the down CURSOR to highlight the next Rate and
Volume fields.
Fe
2
1000
2000
5. Use the DATA keys to enter the rate and volume.
3
750
ZERO
6.
Repeat steps 4 and 5 until all of the infusion informa-
ACCEPT
TOT = 350.0 ml
tion is entered.
-
NoTE: If less than 9 steps of infusion information entered,
the last line of data will always be zero. The total volume
infused will be displayed.
7.
Use the CURSOR to highlight "ACCEPT
TOT=," then press ENTER.
8.
Press PRIMARY RUN.
NOTE: The pump will begin infusing the Period 1 volume at
the Period 1 rate. The pump will then go to Period 2 and so
forth without user intervention.
NOTE: ACCEPT cannot be selected when 0 data is present
in the middle of a program. For example, Pd. 1 and 3 may
have data, but Pd. 2 is zeros. The pump will not allow
ACCEPT to be selected with zero values in the middle of
the program.