OUTLOOK" 200 wir Dosescan"" OPERATOR'S MANUAL
EPIDURAL IN DOSE MODE
When the Epidural category from the Drug List is selected,
-
the following screen will appear:
Pressure will be set to 750
mmHg, and MicroRate will be
enabled.
Use Seis labeled for Epidural
use only.
CauTion: Whenever a pump is used for epidural infusion,
the pump's secondary features should not be used.
CAUTION: To prevent inadvertent infusion of drugs which
are not indicated for epidural use, only Horizon® micro-
bore administration sets without injection sites are
specified for epidural administration of drugs.
WARNING: Serious injury to the patient may result from
epidural administration of drugs other than those
specifically labeled for epidural use.
DOSEGUARD™
IN DOSE MODE
The Outlook™ 200 is equipped with a special feature
which alerts the caregiver when a dosed infusion is being
set outside the infusion parameters specified by the insti- | This value is outside of the
tution. When the user attempts to validate the Data Entry | allowable dosage limits?
Screen, the pump will compare the entered dosage, height,
Continue?
and weight values against the infusion parameter minimum
YES
and maximum. If the values are outside of the set
parameters, a pop-up screen will appear with a message:
After checking that the entered parameters are correct, the caregiver may then select "YES" to
confirm the data and proceed with the infusion.
NOTE:
If either dosage or concentration units have been
changed in the Set Up Screen (see "TO CHANGE UNITS ON
THE DATA ENTRY SCREEN"), and are not compatible with the
currrent dosing units, then Drug Name and DoseGuard™
will be disabled.
34
950952 Rev A (9/03)