Page 1
A732-01-880 Issue D Instruction Manual XDS Dry Pump Description Electrical Supply Item Number XDS100B Scroll Booster 100-120 V, 200-230 V, 50/60 Hz - Set to 200-230 V A732-01-983 Original Instructions...
Page 2
Declaration of Conformity Edwards Ltd, Innovation Drive, Burgess Hill, West Sussex, RH15 9TW, UK The following products: XDS35i scroll pump, 100‐120 V, 200‐230 V, 1 ph, 50/60 Hz, set to high volts A730‐YY‐983 XDS35i scroll pump, 100‐120 V, 200‐230 V, 1 ph, 50/60 Hz, set to low volts A730‐YY‐986 XDS35i scroll pump, 100‐120 V, 200‐230 V, 1 ph, 50/60 Hz A730‐01‐983 XDS35i scroll pump no gas ballast, 100‐120 V, 200‐230 V, 1 ph, 50/60 Hz A730‐05‐983 XDS35iC scroll pump, 100‐120 V, 200‐230 V, 1 ph, 50/60 Hz A730‐06‐983 XDS35iE scroll pump enhanced, 100‐120 V, 200‐230 V, 1 ph, 50/60 Hz A730‐03‐983 XDS35iE scroll pump no gas ballast enhanced, 100‐120 V, 200‐230 V, 1 ph, 50/60 Hz A730‐07‐983 XDS35iCE scroll pump enhanced, 100‐120 V, 200‐230 V, 1 ph, 50/60 Hz ...
Page 3
P200-10-019 Issue D Material Declaration In accordance with the requirements of the Chinese regulatory requirement on the Management Methods for the Restriction of the Use of Hazardous Substances in Electrical and Electronic Products Order No. 32 (also known as ‘China RoHS2’) and SJ/T 11364 Marking for the Restricted Use of Hazardous Substances in Electronic and Electrical Products: Product Labels Product...
Page 4
This page has been intentionally left blank.
Introduction Scope of this manual This manual provides installation, operation and maintenance instructions for the Edwards XDS100B Scroll Booster. The pump must be used as specified in this manual. Read this manual before installing and operating the pump. Important safety information is highlighted as WARNING and CAUTION instructions; these instructions must be obeyed.
WARNING Do not expose any part of the human body to vacuum as it can cause injury. WARNING The Edwards XDS100B booster is not recommended for pumping explosive gases or hazardous substances. WARNING Use suitable lifting equipment to move the XDS100B booster. The mass of the pump is approximately 48 kg.
If an exhaust isolation valve is fitted, ensure the pump cannot be operated with the valve closed. Refer to Section 3.6. Dilute flammable mixtures to safe concentrations by providing an inert gas dilution purge. Contact the Edwards applications team for further advice on dilution requirements if required and refer to Section 4.5.2.
Provide a firm, level platform for the pump. Locate the pump so that the Run/Standby switch is accessible. 3.4.1 Mechanical fixing CAUTION Use the four holes located on each corner of the pump base to secure the pump, if required. Edwards recommends the use of M8 bolts. Electrical installation 3.5.1...
Ensure that the sealing surfaces are clean and scratch-free. Edwards recommends using an exhaust extraction system suitable for use with all process gases being pumped. Ensure that the exhaust extraction system cannot become blocked or obstructed when the pump is operating.
Note: Boost operation is not recommended for the XDS100B as a permanent operation mode, as increased bearing loads will reduce the life of the pump bearings. Please consult Edwards if the boost mode is intended to be used. Shut down WARNING When the pump has been switched off the fan will continue to run for 1 minute.
(which may include hydrofluoric acid) if they are heated to 260 °C and above. The pump may have overheated if misused or in a fire. Health and Safety Data sheets for fluorinated materials used in the pump are available on request; contact the supplier or Edwards. Maintenance plan...
Test the earth continuity and the insulation resistance of the pump motor, in accordance with local regulations for the periodic testing of electrical equipment. Edwards recommends that the earth continuity is less than 0.1 and the insulation resistance is greater than 2M. (Reference EN61010-1).
Equipment which has been serviced, repaired or rebuilt is returned with a full warranty. The local Service Centre can also provide Edwards engineers to support on-site maintenance, service or repair of the equipment.