3.5.2. Manual Calibration

For certain applications or under certain circumstances a temperature offset or a change of the gain for the
temperature curve may be useful.
The factory default settings for Offset and Gain are:
▪ Offset:
0,0 K
▪ Gain:
1,000
A changed Offset causes a parallel shifting of the temperature curve and therewith it has a linear effect on the
temperature reading (change constant independent on process temperature). A change of the Gain will have
a non-linear effect on the temperature reading (change depends on process temperature).
-