 Connect the black voltage test lead to the "N" terminal.
 Connect the red voltage test lead to the "U1" terminal.
 Connect the yellow voltage test lead to the "U2" terminal.
 Connect the blue voltage test lead to the "U3" terminal.
 Connect the I1 current probe output plug to the "I1" jack.
 Connect the I2 current probe output plug to the "I2" jack.
 Connect the I3 current probe output plug to the "I3" jack.
 Connect the I4 current probe output plug to the "I4" jack.
4. Connect the voltage test leads and current probe to the electrical
equipment to be tested.
• If possible, before connecting the voltage test leads and current
probe to the electrical equipment to be tested, take off the electrical
equipment's power.
 Connect the black voltage test alligator to the neutral line "N".
 Connect the red voltage test alligator to the line "A".
Connect the yellow voltage test alligator to the line "B".
11
Connect the blue voltage test alligator to the line "C".
12
13
Press I1 current probe trigger to open the jaw and fully enclose the Line
"A".
14
Press I2 current probe trigger to open the jaw and fully enclose the Line
"B".
15
Press I3 current probe trigger to open the jaw and fully enclose the Line
"C".
16
Press I4 current probe trigger to open the jaw and fully enclose neutral line
"N".
5. Press "POWER" key to select (P1, Q1, S1, PF1), (P2, Q2, S2, PF2),
(P3, Q3, S3, PF3) and (Pt, Qt, St, PFt) measured values.
6. Frequency (Hz), Phase Angle (Θ), Neutral Line Current (I4) and Power
Factor (pF) measurement :
 Press "PFΘ" key to cycle shows the PF and Θ measured values.
User Manual