Measurement Considerations
• Center the conductor inside the Current Clamp jaw.
• Make sure the clamp is perpendicular to the conductor.
• For optimal reading, make sure the conductor is positioned between the
alignment marks on the jaws of the Current Clamp.
Observe the following guidelines when making measurements :
• Avoid taking measurements close to other current carrying conductors.
Operation
To use the AC Current Adaptor, refer to "Measurement Considerations" and
Figure 1.
1. Connect the clamp cable to the instrument.
2. The arrow on the top of the clamp must face towards the load of the
circuit.
3. Connect the current clamp jaws around the conductor to be measured.
Warning
To avoid shock or personal injury, keep fingers behind the tactile barrier, see
Figure 1.
User Manual
Figure 1
Load Direction Arrow
Alignment Marks
Tactile Barrier