Periodic Recording - Saadat DENA 1210 User Manual

Electrocardiograph
Hide thumbs Also See for DENA 1210:
Table of Contents

Advertisement

Chapter 4: Recording Operation

Periodic Recording

This mode is used to generate consecutive records with a time interval and specified repetition.
Recording in this mode is always done according to the latest settings of the device and
changing the settings does not interfere with this operation. Also, the number and time
remaining until the next record is always displayed in the header area.
To perform recording operations in periodic mode:
1- First, by entering the Setting menu and selecting the Recording section, go to the
Periodic Recording Interval option and select the desired time intervals for recording
from the 5-60 minutes options. By touching each section, a drop-down list with
different options is provided to the user; By dragging it up and down, the desired value
can be found and by touching it again, that option is selected and saved.
2- Choose the number of times (Periodic Recording Repetition) from options 1 to 20
and.∞.
3- Recording mode in this type, is similar to other types of recording and is determined
using the Rec. Mode menu.
Returning to the main screen, the number and time remaining until the next record is displayed
in the header area of the screen, and when the counter number reaches zero, periodic recording
according to the current settings of the device begins.
Note
• In this case, the recording is always done according to the latest settings of the device.
• If needed, during periodic recording, recording can be done in Auto, Manual and
Rhythm modes. For this purpose:
1- Set the desired settings and Rec Mode.
2- Press the "Start / Stop" key.
At the end of this operation, periodic recording will resume automatically according
to the latest settings.
• It should be noted that if the device is set to Manual but periodic recording is active,
Rec Type will change to the equivalent Auto mode and remains as long as the
recording settings are not changed by the user, periodic recording will be performed
in this mode. For example, it records Manual 3+ like Auto 3+ mode.
• When it is time to record periodically, no recording will be performed while each
menu is open, and the message "Rec's waiting" will be displayed in the header area.
After returning to the main page, periodic recording begins.
• It is possible to copy the stored information only in Auto and Rhythm modes (since
recording in periodic mode is also done in automatic mode, it is also possible to copy
in this mode).
38