Chapter 5: Recording operation
Only in Automatic and Periodic modes, stored data can be copied.
Rhythm Recording
Select Rhythm using the Mode key on the screen or control panel to see ECG waveform of the main
lead in four traces. Press Start/Stop key to record according to "Length of Rhythm Rec" (Refer to
chapter 2 "User Setting Menu").
There are always six channels of recording in this mode.
Periodic Recording
To perform periodic recording:
1. Set On "Periodic Recording" and select your desired time interval (5-60 min).
2. Select number of recording repetitions. Available options are "infinite" and 1-20.
3. Select recording mode using Mode key.
(For more information about Periodic recording settings, refer to "User Setting Menu" in chapter 2).
You can also perform Automatic and Manual recordings during Periodic recording.
For this purpose:
1. Select the recording mode.
2. Press the Start/Stop key.
After the recording in the selected mode is finished, Periodic recording will be started
automatically.
Periodic recording is performed only in automatic modes.