Chapter 2: System settings

User Setting menu

By selecting User Setting from the main menu, the following window will appear:
The following settings are applicable in this menu:
• Pace: Available options are Off and On. DENA 1210 detects and rejects pacemaker-
generated signals from ECG signal, so that they will be ignored in determining heart
rate.
For patients with Pacemaker, a red indicator is displayed on the screen where the Pace
signal is detected in the ECG signal. In the record sheet, vertical bars are drawn at the
location of the pace.
• Save: If this option is turned on, all Auto and Rhythm mode signals are saved along
with patient information and are available in the Archive menu.
• Meas. & Interp.: If you select the Global or Details options, the corresponding
measurement table will be printed at the end of the record. See the Measurement and
Interpretation chapter for more information. This option may not be enabled on your
device at all. In order to activate it, contact the after-sales service unit.
• Header: By selecting On, the desired signal is printed along with the recording
information at the beginning of the recorder paper, and if selected Off, this information
is not printed and only the desired heart signal is printed without any information.
Warning
• For patients with Pacemaker, Pace should be On. Otherwise, Pacemaker signals may
be considered as QRS.
• In patients with Pacemaker, if Pace is Off, it is best to turn off the 25Hz, 35Hz and
EMG filters to diagnose Pacemaker malfunctions.
Figure 2-6 User Setting menu