Rec. Setting Menu - Saadat DENA 1210 User Manual

Electrocardiograph
Hide thumbs Also See for DENA 1210:
Table of Contents

Advertisement

Chapter 2: System settings

Rec. Setting menu

By selecting Rec Setting from the main menu, the following window will appear:
The following settings are applicable in this menu:
• Rec Time (Auto mode only): This option is used to set the recording time of leads in
Auto mode, the available options are 3-12 seconds.
• Paper Speed: Used to set the recording speed. Available options are 6.25, 12.5, 25 and
50 (mm / sec).
• Gain: Used to adjust the ECG amplitude in the main screen and record. Available
options are 2.5, 5, 10, 20 and Auto (mm / mV).
Note
If Smart Record is activated, the Auto option for Gain is also activated and displayed
and can be selected by the user. If this option is disabled, the Auto mode in Gain is
also removed and its default option (10 mm/mV) is selected.
• If you select the Auto option, the device will automatically select the best gain for
recording and display screen, and its value will be entered in the record header.
• Also, if the patient has a pacemaker and the Pace option is OFF, the Auto Gain
function will be impaired.
Figure 2-4 Rec. Setting menu