User Setting Menu - Saadat Dena 640 User Manual

Electrocardiograph
Table of Contents

Advertisement

User Setting Menu

By pressing User Setting in the Main menu, you can access User Setting Menu (Figure 3-9).
The following parameters can be set in this menu:
Rec Mode: Available options are Real Time and Sync in Auto.
In the Sync mode, the signals of different leads are recorded simultaneously i.e. recording of all
leads starts at the same time.
Sync recording can be performed only in Automatic and Periodic
modes.
Rhythm Lead: The Rhythm lead can be one of the leads I, II, III, aVL, aVR, aVF, V1, V2, V3,
V4, V5 and V6.
Low Pass Filter: Press to toggle between 25, 35, 75 and 150 Hz. This filter is used to remove
muscle artifacts and high -frequency noises. The frequency of selected filter is displayed on the
screen.
Rec Time: Press to set recording duration for different leads in Auto mode. It ranges from 3 to 12
seconds.
Length of Rhythm Rec: Press to set duration of Rhythm lead recording. Available options
are 30, 60, 90, 120, 150 and 180 seconds.
Figure 3-9 User Setting Menu