Chapter 3. Device Setting
3-4 User Setting Menu
By pressing User Setting in the Main menu, you can access User Setting Menu (Figure 3-7).
The following parameters can be set in this menu:
Rec Mode: Available options are Real Time and Sync in Auto.
In the Sync mode, the signals of different leads are recorded simultaneously i.e.recording of
all leads starts at the same time.
Sync recording can be performed only in Automatic and Periodic
modes.
Rec Time: Press to set recording duration for different leads in Auto mode.It ranges from
3to12 seconds.
Hospital/Ward: By selecting this option, you can access the following window and enter
hospital or ward name.
Figure 3-7 User Setting Menu
25
.