Promax OD-610 User Manual page 87

Digital oscilloscope
Table of Contents

Advertisement

Figure 80. Set the network parameters of the Oscilloscope Software.
Set the network parameters of the oscilloscope. Since the
oscilloscope cannot support obtaining IP Address automatically, you
should assign a static IP address. In the oscilloscope, press the Utility
button and press H1 button, turn the M knob to select the LAN Set;
press the H2 button, the set menu is displayed on the right. Set the IP
and the Port to the same value as the "Ports-settings" in the software
in step (3). The Netgate and Net mask should be set according to the
router. Press the H3 button to select "Save set", it prompts "reset to
update the config". After resetting the oscilloscope, if you can get data
normally in the oscilloscope software, the connection is successful.
Figure 81. Set the network parameters of the oscilloscope.
82
Page
June 2017