The following functions can not work in the XY Format:
Reference or digital wave form.
Cursor.
Time base control.
Trigger control.
FFT.
Operation steps:
Press the Display button and call out the Display Set Menu.
Press the H3 menu selection button to set XY Mode ON. The display
format is changed to be XY mode (see Figure 49.).
Cymometer
It is a 6-digit cymometer. The cymometer can measure frequencies from
2Hz to the full bandwidth. Only if the measured channel has triggering
signal and in Edge mode, it can measure frequency correctly. In the Single
trigger mode, it is a one channel cymometer and it can only measure the
frequency of the triggering channel. In the ALT trigger mode, it is a two
channel cymometer and it can measure the frequency of two channels. The
cymometer is displayed at the right bottom of the screen.
52
Page
Figure 49. XY Display Mode.