Thermo Scientific Smart-Vue Installation Manual page 16

C02/temperature module
Hide thumbs Also See for Smart-Vue:
Table of Contents

Advertisement

Installation procedure
Thermo Fisher Scientific
5. For chambers being monitored without an access port you may elect
to route the flat ribbon cable through the door with caution. Please
note that passing the flat ribbon cable through the door gasket in
some CO
incubators with an elevated humidity level may cause
2
condensation to form on the door and create other control issues
along with the increased potential for contamination.
Figure 7. Smart-Vue end-point module mounted on incubator door
6. Connect the sensor and wireless module cables by joining the
connectors (being careful to not unscrew them).
Ribbon cable to
sensor unit
Figure 8. Join cable ends by pressing them firmly together
7. Use the provided plastic cable holders to attach or coil the excess
cable neatly.
Note: When routing the cable for the Smart-Vue probe, avoid direct
contact with or close proximity placement of the sensor cabling with any
high voltage wiring. Cabling should be placed with no less than a
minimum of 5 cm (2 inches) distance from high voltage components.
Also, avoid running the probe cable in parallel with high voltage wiring.
For all other manufacturers' equipment, contact the manufacturer of the
instrument for instructions regarding proper placement of the probe.
                                                                            
Ribbon cable to
Smart-Vue
Smart-Vue CO
and Temperature Module Installation Guide
2
13