Appendix
Smart-Vue end-point
module specifications
Thermo Fisher Scientific
•
ISM (Industrial Scientific Medical) band with four frequencies:
US/CAN 915 MHz; Europe 868 MHz; APAC 434 MHz; India and
other countries 867 MHz
•
Channel width: 50 kHz
•
Frequency deflection: 16 KHz
•
Transmission speed: 9600 Baud in NRZ mode
•
Modulation type: GFSK
•
Driven receiver sensitivity for BER= 1%: -107dbm to -110 dBm
•
Driven transmission: 8 dBm to 10 dBm
•
Power output: 25 mW for end-point module operating at 915MHz,
868 MHz, or 867 MHz. For those countries using 434 MHz, the power
output is 10mW (check with your Thermo Fisher Scientific
representative for international product availability).
•
Plastic enclosure: ABS with IP65 protection for indoor use
•
Wireless end-point module operating conditions: 0°C to +50°C (+32°F
to +122°F); 0 to 90% RH, non-condensing
•
Size: 132.7 x 64.2 x 34 mm (5.2 x 2.5 x 1.3 in.)
•
Weight: 150 g (5.3 oz.)
Smart-Vue CO
and Temperature Module Installation Guide
2
23