Placing The Smart-Vue Module - Thermo Scientific Smart-Vue Installation Manual

C02/temperature module
Hide thumbs Also See for Smart-Vue:
Table of Contents

Advertisement

Placing the sensor and
Smart-Vue end-point
12
Smart-Vue CO
and Temperature Module Installation Guide
2
1. After ensuring that the dual CO
disconnected from the module wire connector, place the sensor
inside the chamber to be monitered. Route the sensor connector and
module
flat ribbon cable through an available access port in your chamber.
When returning the access port plug, stopper, or filter, make sure
that the flat ribbon cable to the CO
or flush with the surface of the chamber wall and the access port
plug, stopper, or filter.
2. Avoid placing the CO
in contact with the sensor (e.g. a humidity pan or active humidity
source).
3. After routing the CO
cable outside of the chamber being monitored, connect the 4-pin
male end connector on the sensor end to the 4-pin female connector
on the module end.
Figure 6. Ribbon cable lies flat beneath door joint
4. Mount the Smart-Vue end-point module on the exterior of the
chamber being monitored or on a suitable nearby surface (e.g. the
wall) using the provided plastic holder, which you may mount using
a magnet or Velcro
for your situation, and place the Smart-Vue end-point module as
described on page 9.
/temperature sensor is
2
/Template sensor is installed flat
2
sensor near water which may accidently come
2
/Temperature sensor connector and ribbon
2
®
. Choose the method that is most appropriate
Installation procedure
Thermo Fisher Scientific