Requirements and
recommendations
8
Smart-Vue CO
and Temperature Module Installation Guide
2
For the purposes of this installation guide we assume that:
•
Smart-Vue Server/Smart-Vue Client software is installed and configured
on your computer, and you have appropriate access rights to configure
devices.
•
A Thermo Scientific receiver is installed and functioning on your
computer or network.
•
The personal computer (PC) running Smart-Vue Server software runs
continuously, 24/7/365, and should be connected to an Uninterruptible
Power Supply (UPS) that protects against power surges and provides
power to back up the PC and components.
•
The AC adapter used for your serial receiver (if applicable) should also
be plugged into a UPS.
•
A UPS is also recommended as a power back-up for your
communication/network systems.
•
Weekly manual system testing (e.g., temporarily remove a sensor to
verify the communications system is working) should be performed as
defined in your Standard Operating Procedure (SOP).
•
Recommended maintenance procedures should be followed.
•
If you are storing cold products, Thermo Fisher Scientific recommends
use of a back-up cooling system (e.g., CO
chamber temperature below the critical level should a power failure
occur. Contact your local sales representative for more information.
•
Thermo Fisher Scientific recommends Installation and Operational
Qualifications (IQ/OQ) be performed before initial use. Contact your
local sales representative for more information.
•
For timely notification of an emergency/alarm, it is critical to establish
primary, secondary and tertiary call-out procedures with escalation so
contact is not dependent on reaching one individual. Ideally, ultimate
contact is with a security department or company with 24/7
monitoring.
End-point modules with internal probes should not be placed in
cold-rooms.
Please see the relevant product manuals for details on installing Smart-Vue
receivers and Smart-Vue Server/Smart-Vue Client software.
What is a Smart-Vue wireless end-point module?
or LN
) to maintain freezer
2
2
Thermo Fisher Scientific