CLEANING & STERILIZATION (continued)
Rubber Goods
Follow hospital procedures and sterilizer
manufacturer's
instructions
sterilization of hoses, breathing bags, and
other components of the breathing system.
After sterilization with ethylene oxide (cold
cycle), properly aerate rubber goods.
Inspiratory and Expiratory Valves
Refer to the detailed instructions in the
Absorber System Operator's Instruction
Manual.
70
RETURN TO CD-ROM TABLE OF CONTENTS
Manual Sphygmomanometer
I n
for
the
Sphygmomanometer will not require any
further cleaning than a wipe down with a
liquid disinfection agent. However, if further
disinfection
sphygmomanometer gauge assembly, hoses,
and blood pressure cuff from the anesthesia
machine and sterilize them with ethylene
oxide gas (cold cycle), followed by appropriate
aeration as per the sterilizer manufacturer's
instructions.
NOTE:
withstand the heat of autoclaving, and
therefore must NOT be autoclaved.