Dräger NARKOMED 3 Operator's Instruction Manual page 94

Anesthesia system
Hide thumbs Also See for NARKOMED 3:
Table of Contents

Advertisement

APPENDIX 2: NARKOMED 3 ALARM MESSAGES (continued)
CLASS
ALARM MESSAGE
C
% ENF LOW
C
% ISO LOW
C
AGT DETECTED
A
O2 SENS DISC
A
OXI SEN DISC
A
INS CO2 HIGH
A
% OXYGEN HI
A
AC PWR FAIL
A
REVERSE FLOW
A
VOL SEN DISC
A
O2 CAL DUE
A
O2 NOT CAL
A
O2 SENS ERR
86
RETURN TO CD-ROM TABLE OF CONTENTS
CONDITION
Enflurane % < low limit
Isoflurane % < low limit
Gas monitor agent detected, none
selected
Sensor cord disconnected
Sensor cord disconnected
Inspiratory CO
> 5 mm Hg
2
% O
> high limit
2
NM3 not receiving line power
Reverse flow > 20 ml
Sensor cord disconnected
18 hrs since last O
analyzer
2
calibration
O
analyzer not calibrated
2
O
analyzer sensor error
2
AUDIO
INTERMIT
INTERMIT
INTERMIT
SINGLE TONE
SINGLE TONE
SINGLE TONE
SINGLE TONE
SINGLE TONE
SINGLE TONE
NONE
NONE
NONE
NONE

Advertisement

Table of Contents
loading

Table of Contents