Av Sequential Or Ddd Pacemakers - Dräger Infinity Vista Instructions For Use Manual

Table of Contents

Advertisement

WA R N I N G : F u s i o n m o d e pa c e r d e t e c t i o n i s n o t i n t e n d e d f o r
u s e w it h l a r g e - v o l ta g e , u n i p o l a r pac e m a k e r s . I t i s i n t e n d e d
f o r u s e o n l y w i t h b i p h a s i c pa c e m a k e r s . P l e a s e o b s e r v e t h e
f o l l o w in g :
z
B e f o r e s e l e c t i n g P a c e r F u s i o n M o d e , b e c e r ta i n t h a t t h e
pa t i e n t h a s a b i p h a s i c pa c e m a k e r ( e x t e r n a l o r
i m p l a n t e d ) a n d t h a t i t i s a c c u r a t e l y p r o g r a m m e d a s
a p p r o p r i a t e f o r t h a t pa t i e n t .
z
D o n o t s e l e c t F u s i o n M o d e i f y o u a r e u n c e r ta i n w h a t
t y p e o f pa c e m a k e r i s i n u s e o r h o w i t i s p r o g r a m m e d .
z
S e l e c t F u s i o n M o d e o n l y t o s u p p r e s s r e p e a t e d f a l s e
a s y s t o l e a n d / o r f a l s e l o w h e a r t r a t e a l a r m s .
z
S e l e c t i n g P a c e r F u s i o n M o d e m a y i n c r e a s e t h e r i s k o f
t h e m o n i t o r f a l s e l y c o u n t i n g pa c e r s p i k e s a s Q R S
c o m p l e x e s a n d m a y c a u s e c a r d i a c a r r e s t t o g o
u n d e t e c t e d . T h e r e f o r e , s p e c i a l s u r v e i l l a n c e o f a n y
pa c e m a k e r pat i e n t m o n i t o r e d w i t h t h i s m o d e i s s t r o n g l y
r e c o m m e n d e d .
WA R N I N G : T h e m o n i t o r m a y n o t c o u n t h e a r t r a t e s a c c u r a t e l y
a n d m a y m i s i n t e r p r e t r a t e - d e p e n d e n t a r r h y t h m i a i n s o m e
pa c e d pa t ie n ts . D o n o t r e l y e n t i r e l y o n t h e d i s p l a y e d h e a r t
r a t e t o a s s e s s a pa c e d pa t i e n t 's c o n d i t i o n . A l w a y s o b s e r v e
t h e s e pa t i e n ts c l o s e l y a n d m o n i t o r a l l o f t h e i r v i ta l s i g n s
c a r e f u l l y.

AV Sequential or DDD Pacemakers

For pacemaker patients, follow these steps:
z
Always attach at least four electrodes to allow for a choice of leads.
z
Turn pacer detection on.
z
Select the lead with the highest R-wave and the least interference.
z
Verify that the HR calculation is accurate by comparing it to the ECG
waveform.
z
Monitor the pulse or the respiration or both using a method other than ECG.
For patients without pacemakers, turn the pacer detection off.
VF6
ECG
HR S
AND
Infinity Vista
C
AFETY
ONSIDERATIONS
8-19