NARKOMED 3 SETUP & INSTALLATION (continued)
8.
Open the cylinder valve by turning
the cylinder valve stem in a counter-
clockwise direction, using the cylinder
wrench tethered to the rear of the
NARKOMED 3.
9.
Check for adequate cylinder pressure
on the pressure gauge on the
anesthesia machine.
10.
Turn the System Power switch to
"ON" and verify the gas flow from the
exchanged cylinder by opening the
corresponding flow control valve and
observing its flowmeter.
11.
If the oxygen cylinder was exchanged,
verify the delivery of pure oxygen to
the patient breathing system by using
the oxygen analyzer.
12.
Close the flow control valve(s) and
turn the System Power switch to the
"STANDBY" position.
Electrical Power Connection
The NARKOMED 3 is equipped with a power
cable with a hospital grade plug for the 117 VAC
primary electrical power or with a power cable
with a hospital grade plug for the optional 240
VAC primary electrical power. Attach the
NARKOMED 3 to electrical power as follows:
56
RETURN TO CD-ROM TABLE OF CONTENTS
1.
Turn the NARKOMED 3 System Power
switch to "STANDBY."
2.
Unwrap sufficient length of power cord.
3.
Plug the power cable into a 117 or 240
VAC hospital grade outlet (depending on
the voltage requirement of the machine).
4.
Turn the System Power switch of the
NARKOMED 3 to "ON." The central
monitoring display should now be
active.
5.
Verify that the battery-in-use ("AC
POWER FAIL") indicator on the main
switch panel remains extinguished.
6.
Verify that all circuit breakers, located
on the lower right side of the frame,
remain reset (i.e., pushed in).
7.
Turn the System Power switch of the
NARKOMED 3 to "STANDBY."