NARKOMED 3 SETUP & INSTALLATION (continued)
Absorber System and Monitor Sensors
Install the absorber system and monitor
sensors as described below (Figs. 34 & 35).
This installation presumes the use of an
appropriate scavenger system with the
absorber system. If using a Bain circuit
adapter, consult the manual supplied with the
adapter.
1.
Slip the absorber mounting stud into
the absorber pole on the swivel arm.
2.
Tighten the set screw on the absorber
pole to lock the absorber in place.
3.
Pull the fresh gas locking bar, located
on the front of the NARKOMED 3,
out to its extended position (Figs. 11
& 34). Attach the fitting on the fresh
gas hose to the terminal. Release the
spring-loaded locking bar over the
fitting, allowing it to "lock" the fitting
into place.
4.
Attach a 22 mm breathing hose (Fig.
35) between the ventilator bellows 22
mm terminal marked "VENTILATOR
HOSE" and the 22 mm terminal on
the rear of the manual/automatic
selector valve, which is also marked
"VENTILATOR HOSE."
5.
Attach a 22 mm breathing hose (Fig.
34) between the 22 mm hose terminal
on the inspiratory valve marked
"INSPIRATION" and one side of the
Y-piece.
6.
Similarly, attach another 22 mm
breathing hose (Fig. 34) between the
other side of the Y-piece and the 22
mm hose terminal on the expiratory
valve marked "EXPIRATION."
58
RETURN TO CD-ROM TABLE OF CONTENTS
7.
Attach the breathing bag to the swivel
bag mount 22 mm terminal marked
"BREATHING BAG" (Fig. 34).
8.
Connect the 19 mm scavenger hose
between the 19 mm terminal (marked
"SCAVENGER HOSE") on the bottom
of the absorber pole (Fig. 34) and the
1 9 m m
t e r m i n a l
"SCAVENGER
scavenger (Fig. 35).
9.
Attach the breathing pressure pilot line
to the connector on the system interface
panel
marked
PRESSURE" (Fig. 35). If the shorter
pilot line is used, it reads the pressure
at the absorber top dome gas pipe by
means of a quick-connect fitting. If the
longer pilot line is used, it reads the
pressure at or near the patient Y-piece,
at either a 15 mm adapter with the
appropriate Luer fitting or a Y-piece
with the appropriate Luer fitting.
10.
Connect the Oxygen Analyzer sensor
cord to the connector on the interface
panel marked "OXYGEN SENSOR"
(Fig. 35). Make sure that the oxygen
sensor assembly is properly mounted in
the inspiratory valve dome.
11.
Connect the 19 mm scavenger hose
between the 19 mm terminal (marked
"SCAVENGER HOSE") on the rear of
the APL valve and the 19 mm terminal
(marked "SCAVENGER HOSE") on the
rear of the absorber pole (Fig. 34).
12.
Install the Respiratory Volume Monitor
sensor between the expiratory valve
and the absorber. Plug the sensor cord
into the connector on the interface
panel marked "VOLUME SENSOR"
(Fig. 35).
( m a r k e d
HOSE")
on
the
"BREATHING