Chapter 9 Serial Interfaces
9.8.1.10 :STAT:PRESet
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.1.11 :STAT:QUES:ENABle[?]
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.1.12 :STAT:QUES:EVENt?
Description:
Mode:
Parameter:
Response:
*RST
Example:
Page 63 / 92
Reset the measurement event register and the standard event register.
All relevant bits are set to 0.
Command
none
none
Value:
not relevant
send
Set or query the error data event release register. The gaussmeter
returns the bit pattern as a decimal value.
Command and query
{<Value>} within the range of 0..255
Error event release register
Value:
not relevant
send
send
receive
Query the error event register. The gaussmeter returns the bit pattern
as a decimal value.
Query
none
Error event register
Value:
not relevant
send
receive
:STAT:PRES<LF>
:STAT:QUES:ENAB 64<LF>
:STAT:QUES:ENAB?<LF>
64<CR><LF>
:STAT:QUES:EVEN?<LF>
0<CR><LF>