on the selected measuring unit. The ranges can only be switched manually. Normally,
the measuring range is adjusted in the set-up menu. Regarding the manual switch,
the measuring range is selected by multiple pressing of a button assigned in the set-
up menu.
The following table shows an overview of the measuring ranges.
Unit
Range
Tesla
1
T
2
3
4
Ampere/m
1
A/m
2
3
4
Gauss
1
G
2
3
4
Oersted
1
Oe
2
3
4
The values in [] apply to the operation with high-field probes.
Page 26 / 92
DC Field Measurement
Range Limit
Resolution
10 mT
1 µT
100 mT
10 µT
1000 mT
100 µT
4500 mT
1 mT
[10 T]
10 kA/m
1 A/m
100 kA/m
10 A/m
1000 kA/m
100 A/m
3500 kA/m
1 kA/m
[8000 kA/m]
100 G
10 mG
1000 G
100 mG
10 kG
1 G
45 kG
10 G
[100 kG]
100 Oe
10 mOe
1000 Oe
100 mOe
10 kOe
1 Oe
45 kOe
10 Oe
[100 kOe]
Figure 14 Measuring Ranges
Gaussmeter IGM11 Operating Instructions
AC Field Measurement
Range Limit
Resolution
10 mT
10 µT
100 mT
100 µT
1000 mT
1 mT
3000 mT
10 mT
[6.5 T]
10 kA/m
10 A/m
100 kA/m
100 A/m
1000 kA/m
1 kA/m
2500 kA/m
10 kA/m
[5600 kA/m]
100 G
100 mG
1000 G
1 G
10 kG
10 G
30 kG
100 G
[70 kG]
100 Oe
100 mOe
1000 Oe
1 Oe
10 kOe
10 Oe
30 kOe
100 Oe
[70 kOe]
Chapter 5 Operation
Fast Pulse Measurement
Range Limit
Resolution
10 mT
10 µT
100 mT
100 µT
1000 mT
1 mT
4500 mT
10 mT
[10 T]
10 kA/m
10 A/m
100 kA/m
100 A/m
1000 kA/m
1 kA/m
3500 kA/m
10 kA/m
[8000 kA/m]
100 G
100 mG
1000 G
1 G
10 kG
10 G
45 kG
100 G
[100 kG]
100 Oe
100 mOe
1000 Oe
1 Oe
10 kOe
10 Oe
45 kOe
100 Oe
[100 kOe]