Set or query the upper limit value for the limit value check of each
output.
x represents outputs 1 to 4.
Only relevant for the selected compare operation.
Command or query
{<Value>} within the range of -999999.999 .. +999999.999
The significance is determined by the selected measuring unit.
Measuring unit
Tesla
A/m
Gauss
Oersted
like parameter
Value:
not relevant
send
:DGIO:SET:OUT2:HIGH 1234.56<LF>
send
:DGIO:SET:OUT2:HIGH?<LF>
receive
1234.56<CR><LF>
Set or query the function of the upper button (ENTER) on the front
panel.
Command or query
{
NO|RANG|PKNL|NULL
Parameter
NO
RANG
PKNL
NULL
parameter
like
Value:
not relevant
send
:KEYS:TOP:FUNC RANG<LF>
send
:KEYS:TOP:FUNC?<LF>
receive
RANG<CR><LF>
Gaussmeter IGM11 Operating Instructions
:HIGH[?]
{x}
Limit value unit
mT
A/m
G
Oe
}
Function
No function assigned to this key
Range switching
Reset peak values
Zeroing the probe
Chapter 9 Serial Interfaces
Example
Value
123.4 mT
123.4
1.2 kA/m
1200.0
12.3 kG
12300.0
12.3 kOe
12300.0